B
Ba
asse
e d
drriiffttss ((w
wiitth
ho
ou
utt p
pa
attiie
en
ntt w
we
eiig
gh
htt))
1. Check the manual release cable adjustment.
2. Change non-locking manual valve.
3. Change “A” valve.
L
Liitttte
err d
do
oe
ess n
no
ott llo
ow
we
err iin
n tth
he
e p
po
ow
we
erre
ed
d m
mo
od
de
e
Check the power indicator LED.
1. If blinking constant RED, change the battery.
2. If blinking a patterned AMBER short, short, long flash:
a. Check for broken or disconnected wires.
b. Check for 24 VDC at connector (C) on the main cable by the motor while pressing the retract (-) button. If voltage is
present, replace (in order) the solenoid and/or “B” valve. If voltage is not present, go to step c.
c. Check for 24 VDC on electronics assembly pins, 1 blue and 5 orange, on (F) while pressing the retract (-) button. If
voltage is not present, replace the electronics assembly. If voltage is present, replace the wire harness.
3. If the GREEN light turns on, but does not lower the litter, try the other switch. If the other switch works, replace the bad
switch.
L
Liitttte
err d
do
oe
ess n
no
ott e
exxtte
en
nd
d iin
n tth
he
e p
po
ow
we
erre
ed
d m
mo
od
de
e -- cch
he
ecckk m
mo
otto
orr
Check the motor.
1. If the motor runs, but does not raise the cot:
a. Check the manual release cable for too much tension.
b. Lightly tap the manual locking valve.
c. Replace the manual locking valve.
2. If the motor is stalled, replace the “A” valve.
3. If the light is GREEN, but the motor does not run:
a. Check for 24 VDC at connector (E) on the main cable. If voltage is present and the motor does not run, replace the
hydraulic sub-assembly. If voltage is not present, go to step b.
b. Check for 24 VDC on electronics assembly connection (H) (-) lead on black (+) lead on green while pressing the
extend (+) button. If voltage is not present, replace the electronics assembly. If voltage is present, replace the main
cable.
L
Liitttte
err d
do
oe
ess n
no
ott e
exxtte
en
nd
d iin
n p
po
ow
we
erre
ed
d m
mo
od
de
e -- cch
he
ecckk tth
he
e p
po
ow
we
err iin
nd
diicca
atto
orr L
LE
ED
D
Check the power indicator LED.
1. If blinking constant RED, change the battery.
2. If blinking a patterned AMBER short, short, long flash:
a. Check for broken or disconnected wires.
b. Check for 24 VDC at connector (B) on the main cable by the motor while pressing the extend (+) button. If voltage is
present, replace (in order) the solenoid and/or “A” valve. If voltage is not present, go to step c.
c. Check for 24 VDC on the electronics assembly pins, 2 white and 6 red, on (F) while pressing the extend (+) button. If
voltage is not present, replace the electronics assembly. If voltage is present replace the wire harness.
3. If the GREEN light turns on, but does not lower, try the other switch. If the other switch works, replace the bad switch.
6506-109-002 Rev G.1
25
EN
Содержание Power-PRO XT
Страница 2: ......
Страница 6: ......
Страница 57: ...6506 109 002 Rev G 1 51 EN ...
Страница 58: ...EN 52 6506 109 002 Rev G 1 ...
Страница 59: ...6506 109 002 Rev G 1 53 EN ...
Страница 62: ...B Ba as se e a as ss se em mb blly y 6506 001 012 Rev AA Reference only EN 56 6506 109 002 Rev G 1 ...
Страница 91: ...P Po ow we er r a as ss se em mb blly y 6500 002 031 Rev D Reference only 6506 109 002 Rev G 1 85 EN ...
Страница 94: ...F Fo oo ot t e en nd d a as ss se em mb blly y 6500 102 015 Rev AA Reference only EN 88 6506 109 002 Rev G 1 ...
Страница 103: ...6506 109 002 Rev G 1 97 EN ...
Страница 122: ...EN 116 6506 109 002 Rev G 1 ...
Страница 145: ...P Piillllo ow w 6 61 10 00 0 0 04 41 1 0 03 30 0 Rev E Reference only 6506 109 002 Rev G 1 139 EN ...
Страница 147: ...W Wh he ee ell f fr ro on nt t 6 65 50 00 0 0 00 01 1 0 08 86 6 Rev A Reference only 6506 109 002 Rev G 1 141 EN ...
Страница 166: ...T Tr ra an ns sf fe er r f flla at t 6005 001 001 Rev E Reference only EN 160 6506 109 002 Rev G 1 ...
Страница 185: ......
Страница 186: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6506 109 002 Rev G 1 WCR AB 11 2019 07 ...