Annual Inspection
81
EN
13.6
Max. Device Pressure Test
1. Set to High Flow operating mode (Pneumo Sure XL: Set to Bariatric operating
mode).
2. Select the max. nominal gas flow.
3. Connect a manometer
(1)
and an open Veress cannula
(4)
to the insufflation
tube connection
(3)
/adapter
(2)
.
4. Select the max. gas flow.
5. Start insufflation:
Press the
START
function field. The manometer registers a pulsing pressure
increase. When the pressure stabilizes, the manometer registers a maximum
pressure of 55-65 mm Hg (Bariatric: 65-75 mm Hg).
6. Stop insufflation:
Press the
STOP
function field.
13.7
Gas Flow Rate Test
Test setup with open connection, without connected insufflation tube.
• Select a nominal gas flow rate of 15 l/min.
• Start insufflation:
Press the
START
function field.
• Press the gas volume function field (0.0 l must be displayed).
Now start measuring for one minute.
• Stop the insufflation after one minute: Press the
STOP
function field.
The gas consumption should be at least 11-12 l.
Each successfully conducted test must documented with the test log.
mm Hg
(1)
(2)
(3)
(4)
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......