80
Annual Inspection
EN
13.4
Pressure Monitoring Test in High Flow Operating Mode
See also 16 Technical Data, page 89.
1. Set to High Flow operating mode.
2. Select a nominal pressure of 15 mm Hg and a nominal gas flow of 3 l/min.
3. In the
Venting pressure limit
configuration menu, set a venting pressure of
3 mm Hg.
4. Use the syringe to generate a pressure of at least 19 mm Hg, which registers
on the manometer. Start insufflation: Press the
START
function field. An
acoustic warning sound is emitted with a pressure of more than 19 mm Hg
(for 5 seconds) and the display depicts
Overpressure
.
5. Reduce the pressure. The warning ends when the pressure falls below 19 mm
Hg (nominal pressure plus 4 mm Hg). Stop insufflation: Press the
STOP
func-
tion field.
6. Select a nominal pressure of 29 mm Hg.
7. Use the syringe to generate a pressure of at least 30 mm Hg, which registers
on the manometer. Start insufflation: Press the
START
function field. An
acoustic warning sound is emitted without delay in case pressure exceeds
30 mm Hg and the display depicts
Overpressure
.
8. Reduce the pressure. The warning ends when the pressure falls below
30 mm Hg. Stop insufflation: Press the
STOP
function field.
13.5
Venting Valve Test
See also 13.3 Testing the Pressure Sensors in High Flow Operating Mode, page 79.
1. In the
Venting System
configuration menu, activate the venting system
(when
Veress insufflation OFF
is set).
2. In the
Venting response time
configuration menu, set a venting time of
3 seconds.
3. In the
Venting pressure limit
configuration menu, set a venting pressure of
3 mm Hg.
4. Select a nominal pressure of 15 mm Hg and a nominal gas flow of 10 l/min.
5. Use the syringe to generate a pressure of at least 18 mm Hg, which registers
on the manometer. Start insufflation. The venting valve is activated and the
display depicts
Venting system active
if pressure exceeds 18 mm Hg (for
3 seconds).
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......