Annual Inspection
79
EN
flation tube.
7. Stop insufflation: Press the
STOP
function field.
The following values are displayed:
Actual pressure 0.0 [mm Hg]
Gas consumption >0.0 [l]
8. Press the gas volume function field.
Gas consumption >0.0 [l]
The basic function test of the device is complete.
13.3
Testing the Pressure Sensors in High Flow Operating Mode
1. Set to High Flow operating mode.
2. Select a nominal gas flow rate of 1.0 l/min.
Do
not
press the
START/STOP
function field.
CAUTION!
Never use the syringe to extract gas from the device.
3. Connect a manometer
(1)
and an air-filled syringe
(4)
to the insufflation
tube connection
(3)
/adapter
(2)
.
4. Use the syringe to generate a pressure of at least 10 mm Hg, which registers
on the manometer.
Actual pressure display: 10 ±2 [mm Hg]
5. Use the syringe to generate a pressure of at least 20 mm Hg, which registers
on the manometer.
Actual pressure display: 20 ±2 [mm Hg]
6. Use the syringe to generate a pressure of at least 30 mm Hg, which registers
on the manometer.
Actual pressure display: 30 ±2 [mm Hg]
mm Hg
(1)
(2)
(3)
(4)
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......