Using and Controlling the PNEUMO SURE High Flow Insufflator in High Flow Mode
29
EN
• Keeping the
º
or
»
function field pressed longer than 1.5 seconds activates
scrolling through the gas flow rates 3, 20, 40 l/min or 40, 20, 3 l/min.
2. The preset values of the Configuration menu can be changed individually (see
chapter 10.1 Configuration menu I, page 60). Select a nominal gas flow be-
tween 1-40 l/min. The preset value is indicated in the nominal gas flow dis-
play. The values for the nominal gas flow refer to a device without connected
tube, filter, or instrument. Tube, filter, and instrument can reduce the gas
flow.
The device monitors the gas flow in two different operating modes depicted in
the status line with the following messages:
•
Veress insufflation
(1-5 l/min)
•
Insufflation: High Flow
(6-40 l/min)
Veress insufflation operating mode:
Veress insufflation
is a gentle type of insufflation so that the actual pressure
does not exceed the preset pressure even in case of small volumes. To minimize
the risks in case of a faulty incision, the manufacturer recommends using
Veress
insufflation
to start a procedure (filling abdomen with CO2).
CAUTION!
Please note that the automatic venting system is not active during Veress insuf-
flation mode.
Insufflation: High Flow
While
Insufflation: High Flow
any pressure loss due to leaks can be quickly equal-
ized. The APC Technology (Advanced Pressure Control) enables raising the actual
pressure gently to the level of the nominal pressure. In case of large volumes, the
actual pressure does not exceed the nominal pressure (see chapter 11 Safety
functions, page 70).
Start the device by pressing the
START
function field.
With a nominal flow setting of < 6 l/min
Veress insufflation
is displayed. Above
5 l/min
Insufflation: High Flow
) is displayed.
Stop the device by pressing the
STOP
function field.
NOTE!
Tube, filter, and instrument can reduce the gas flow.
6.5
Performing the Function Test in High Flow Operating Mode
Preparation
Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
For your own safety and that of your patient, use only original accessories.
WARNING!
The function test must be performed prior to each surgical procedure.
START
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
Mode: High Flow
3
15
Change
Mode
Actual
l/min
0
Liter
0.0
0
GAS
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......