Electrical Safety
PIP-Safety Analyzer Setup for IEC62353 Testing
WARNING
POSSIBLE EQUIPMENT DAMAGE
Do not defibrillate when connected to the safety analyzer!
To perform the Direct Equipment Leakage, Direct Applied Part Leakage and Protective Earth Resistance tests in accordance with IEC
62353, set up the Safety Analyzer as follows:
Press the SETUP button on the Safety Analyzer.
1. Press the SETUP button on the Safety Analyzer.
2. Press the F4/MORE button.
3. Press the F2/INSTRUMENT button.
4. Press the F1/STANDARD button.
5. Press the UP/DOWN arrows to select the 62353 Standard.
6. Press the DONE button.