G H
FFiigguurree 1111 –– FFoooott bbooxx ccoovveerr
15.Reverse to reinstall.
SSoolleennooiidd vvaallvvee rreeppllaacceem
meenntt
TToooollss rreeqquuiirreedd::
• Zipper pull tool or equivalent
• Torque driver
• ESD system
PPrroocceedduurree::
1. See (
2. See
Protecting against electrostatic discharge (ESD)
3. Using the supplied torque driver, remove the two screws that secure the solenoid valve (H, J, or K) to the manifold
assembly (Figure 12). Save the screws.
NNoottee -- During installation tighten the screws equally. Use the supplied torque driver and torque the screws to 4.4 in-lb.
2972-009-002 Rev AA.0
33
EN
Содержание Isolibrium 2972
Страница 2: ......
Страница 22: ...I It te em m C Co ollo or r 1 Black 2 Red 4 Green 6 open EN 18 2972 009 002 Rev AA 0 ...
Страница 54: ...EN 50 2972 009 002 Rev AA 0 ...
Страница 55: ...2972 009 002 Rev AA 0 51 EN ...
Страница 56: ...EN 52 2972 009 002 Rev AA 0 ...
Страница 57: ...2972 009 002 Rev AA 0 53 EN ...
Страница 58: ...EN 54 2972 009 002 Rev AA 0 ...
Страница 62: ...Torque item B to 22 4 in lb Torque item B to 22 4 in lb EN 58 2972 009 002 Rev AA 0 ...
Страница 63: ...Torque item A to 22 4 in lb Tighten item BG force 25 40 lb 2972 009 002 Rev AA 0 59 EN ...
Страница 65: ...Torque item C to 8 2 in lb 2972 009 002 Rev AA 0 61 EN ...
Страница 66: ...Torque item B to 22 4 in lb EN 62 2972 009 002 Rev AA 0 ...
Страница 75: ...2 29 97 71 1 0 02 22 2 0 00 08 8 Rev AB Reference only AB 2971 021 033 1 2972 009 002 Rev AA 0 71 EN ...
Страница 76: ...2 29 97 71 1 0 02 22 2 0 00 08 8 Rev AB Reference only AJ 297300220014 1 EN 72 2972 009 002 Rev AA 0 ...
Страница 78: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 2972 009 002 Rev AA 0 WCR AA 5 2021 07 ...