Preventative Maintenance
WARNING
Service only by qualified personnel. Refer to the maintenance manual.
Ensure the power cord is unplugged and battery power switch (Epic+ option) is turned to the off position before
servicing.
NURSE CALL BATTERY
To prevent a low battery condition when the bed is not plugged in, position the cord out switch at the head
end of the bed to the off position. The switch is identified by the label shown below. If the switch is not posi-
tioned as shown below and the bed power cord and pendant cord are unplugged, the life of the back–up bat-
tery will be significantly reduced.
If the power light (located on the foot board) is flashing, the Nurse Call battery needs to be replaced. The
battery is located on the patient’s left side at the head end of the bed. No tools are required to replace the
battery. Unplug the bed power cord from the wall socket and replace the battery.
MAIN BED POWER CIRCUIT BREAKER
In the event of a loss of bed functions, unplug the bed power cord from the wall socket and reset the circuit
breaker(s) located under the bed on the patient’s left side. Plug the bed into a properly grounded wall recep-
tacle and follow the set–up procedures listed on page 6.
BATTERY CHARGER CIRCUIT BREAKER (EPIC+ OPTION)
If the battery charger circuit breaker(s) located under the litter on the patient’s head end, left side are tripped,
refer to the troubleshooting section of the maintenance manual.