42
8001-009-001 REV G
www.stryker.com
Disinfecting
Disinfecting the internal water circuit and hoses (Continued)
5.
Disconnect the bottom hose from the bottom right port (
).
Figure 29: Disconnected hose
6.
Connect the bottom hose end to the hydraulic connector in the lid of the reservoir (
Figure 30: Bottom hose end in the lid of the reservoir
7.
Plug the power cord into a wall outlet.
8.
Press and hold the
Stand-by button.
9.
Tap the Manual mode icon.
10. Tap Confirm.
11. Set the water target temperature to 25.0° C (77.0° F).
12. Tap Confirm.
13. Run the controller for 20 minutes.
1
14. After 20 minutes, turn the controller off by pressing and holding the Stand-by button for two seconds.
15. Unplug the power cord from the wall outlet.
16. Place the controller over a floor drain.
17. Remove the reservoir. Pull forward at an angle, and lift out the reservoir.
18. Remove the bottom hose end from the hydraulic connector adapter in the reservoir lid by pushing down on the collar.
19. Empty water from the reservoir, dispose of the water per hospital protocol.