Operation
Storing the power cord and hoses
After you complete therapy or when you transport the product, store the power cord and hoses.
CAUTION
•
Do not hang items on the controller handle to avoid the risk of tipping the product.
•
Always store the power cord, cables, and hoses before you transport the product to reduce the risk of trip hazard.
To store the power cord and hoses:
1.
Connect the ends of the connector hoses together, if applicable.
2.
Coil and fasten the hose with the management straps (
).
3.
Unplug the power cord from the wall outlet and store with the management straps (
Figure 22: Management straps
Storing the controller
Storage is equal to or greater than 7 days without use.
CAUTION
•
Do not store the product with water in the device.
•
Always store the product within the specified environmental condition values.
To store the controller:
1.
See
Disinfect the internal water circuit and hoses every 14 days on page 39
2.
See
Draining the thermal transfer devices on page 33
3.
See
Cleaning the external surfaces on page 37
.
4.
See
Disinfecting external surfaces on page 38
Notes
•
Always bring the controller to room temperature after high or low temperature storage.
•
Always store the controller with the reservoir in place.