30
8001-009-001 REV G
www.stryker.com
Operation
Setting or editing the warming rates
Setting warming rates is for Automatic Mode only.
1.
To set the warming temperature, highlight your choice of warming rates.
Select a warming rate
Description
M
ax
Maximum: approaches patient target temperature as fast as possible
M
ed
Medium: patient temperature increases at a rate of 4.0° C in 12 hours (0.33° C/ hour).
Min
Minimum: patient temperature increases at a rate of 4.0° C in 24 hours (0.17° C/hour).
S
et Custo m
Set Custom: patient temperature increases at a customized temperature and time
period the operator selects. The temperature increases 0.05° C/hour to 0.5° C/hour.
2.
If you select Set Custom, tap the Increase and Decrease buttons to set the rate (
).
rr
Figure 17: Set custom warming rate
3.
Tap Confirm.
4.
Tap the Edit button to make changes.
Starting Manual mode
In Manual mode, the therapy will cool or warm the water to a selected water target temperature. The operator must
observe the patient’s temperature and manually adjust the water temperature to obtain the desired patient temperature.
1.
If desired, select and place the sensing end of the patient probe based on hospital protocol. Connect the reusable
adapter cable to port A or port B on the product. See
Selecting and connecting a temperature probe on page 22
2.
Prepare the thermal transfer devices to be used for therapy.
3.
See
Filling a thermal transfer device on page 28
4.
Apply the thermal transfer device to the patient.
5.
Tap Manual mode. The default water target temperature is 40.0° C (104.0° F) upon initial entry.
6.
Tap Confirm.
7.
To select the desired water temperature, tap the Increase or Decrease buttons or hold the button to go faster.
a.
To edit water temperature, tap the Edit button.
8.
Tap Confirm.