6400-099-700 Rev-AA

EN

www.stryker.com

11

Specifications

WARNING: 

ALWAYS check any documentation that 

accompanies attachments and/or cutting accessories for 
special duty cycle and usage instructions.

Model:

RemB Electric Universal Driver  
(REF 6400-099-000)

Dimensions:

5.3 inch [136 mm] height 
1.02 inch [26 mm] width 
5.47 inch [139 mm] length

Mass:

1.18 lb [0.53 kg]

Speed:

1500 rpm

Mode of Operation: 

Duty Cycle: 
Rest Between Cycles:

Non-continuous Operation 
20 seconds on/20 seconds off, 6 times 
45 minutes

Power Supply:

Stryker CORE 2, CORE, or TPS 
Console 
40 V 

 (Direct Current)

Equipment Type:

Type BF Applied Part

Maximum 
Temperature of 
Applied Part:

Less than 124 °F [51 °C] (Maximum 
surface temperature as tested to the 
standards listed under 

Product Safety 

Certification

 in the instructions for use 

supplied with the console.)

Ingress Protection:

IPX0 Ordinary Equipment

Environmental 
Conditions:

Operation

Storage and 

Transportation

Temperature 
Limitation:

Humidity 
Limitation:

Atmospheric 
Pressure 
Limitation:

Содержание 6400-099-000

Страница 1: ...2018 10 6400 099 700 Rev AA www stryker com Instructions For Use RemB Electric Universal Driver 6400 099 000 ENGLISH EN...

Страница 2: ...t product damage A NOTE supplements and or clarifies procedural information For additional information especially safety information or in service training contact your Stryker sales representative or...

Страница 3: ...fore use See the Specifications section Upon initial receipt and before each use clean and sterilize the equipment as indicated See the care instructions manual supplied with the equipment Upon initia...

Страница 4: ...acement and or bone or tissue damage If wobbling occurs see the Troubleshooting section When using the Universal Driver always use Stryker attachments with two J slots The Universal Driver will not fu...

Страница 5: ...fety Certification standards listed in the Specifications section of the instructions for use supplied with the console D Attachment E Reverse Trigger Press this pressure sensitive trigger to obtain v...

Страница 6: ...ctional FORWARD REVERSE Fully extend the switch to unlock both triggers Both triggers are functional Definitions The symbols located on the equipment and or labeling are defined in this section or in...

Страница 7: ...e is idle when passing the handpiece to another person or before inserting or removing a cutting accessory or attachment DO NOT press the pedal of a connected footswitch while the function switch is s...

Страница 8: ...en do not function Use the handpiece triggers or a footswitch to select the handpiece direction When using the console the Oscillate button will initiate a turn based oscillate mode that defaults to a...

Страница 9: ...trical malfunction exists Return the handpiece for repair The handpiece will not operate The function switch is in the SAFETY position Set the function switch to the FORWARD or FORWARD REVERSE positio...

Страница 10: ...rference is experienced during installation The release button mechanism needs lubrication See the care instructions manual supplied with the equipment Sporadic electrical interference is experienced...

Страница 11: ...Cycle Rest Between Cycles Non continuous Operation 20 seconds on 20 seconds off 6 times 45 minutes Power Supply Stryker CORE 2 CORE or TPS Console 40 V Direct Current Equipment Type Type BF Applied Pa...

Страница 12: ...700 Rev AA www stryker com Stryker Instruments 4100 E Milham Kalamazoo Michigan USA 49001 1 269 323 7700 1 800 253 3210 ES DE FR IT NL 6400 099 713 JA ZH KO 6400 099 720 SV DA FI PT NO 6400 099 730 PL...

Отзывы: