background image

www.stryker.com

2920-109-001 REV G

27

Disinfecting

WARNING

Always disinfect the support surface between patients, to avoid the risk of cross-contamination and infection.

Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning.
Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly. If
you do not properly rinse and dry the product, a corrosive residue may be left on the surface of the product that
could cause premature degradation of critical components. Failure to follow these cleaning instructions may void
your warranty.

Do not allow fluid to pool on the support surface or foot box. Fluids may cause degradation of components and may
cause unpredictable safety and performance of this product.

Always unplug the product power cord before cleaning or disinfecting to avoid the risk of shock.

CAUTION

Always completely dry the support surface covers before storing, adding linens, or placing a patient on the surface.
Failure to remove excess disinfectant may degrade the cover material.

Do not over expose the covers to higher concentration disinfectant solutions as these may degrade the covers.

Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the
cover.

Prerequisite: Minimum of two operators are required to disinfect the support surface.

Suggested Disinfectants:

Quaternaries

Phenolic Disinfectants

Chlorinated Bleach Solution (5.25% sodium hypochlorite at 1:100 dilution)

70% Isopropyl Alcohol

To disinfect the support surface covers after each patient use, follow these steps in order:

1.

Unplug the support surface.

2.

Thoroughly clean and dry the support surface covers (see

Cleaning on page 25

) before disinfectants are applied.

3.

Apply recommended disinfectant solution by spray or pre-soaked wipes (do not soak the support surface).

Note:

Make sure that you follow the disinfectant’s instructions for appropriate contact time and rinsing

requirements.

4.

Wipe the support surface covers with a clean, dry cloth to remove any excess liquid or disinfectant.

5.

Allow the support surface covers to dry completely before returning to service. Air dry, if possible.

Содержание 2920

Страница 1: ...2016 10 G 2 2920 109 001 REV G www stryker com Operations Manual PositionPRO Patient Repositioning Support Surface Standalone with pendant or integrated with FL27 InTouch CC Model Beds 2920 ...

Страница 2: ...sample text ...

Страница 3: ...PRO is considered an active therapeutic device intended to be used to administer or withdraw energy to or from the body Medical Equipment Recognized by Underwriters Laboratories LLC With Respect to Electric Shock Fire and Mechanical Hazards only in accordance with UL 60601 1 2003 and CAN CSA C22 2 No 601 1 M90 updates 1 and 2 In accordance with European Directive 2012 19 EU on Waste Electrical and...

Страница 4: ...2920 109 001 REV G www stryker com Symbols Chlorinated bleach For US Patents see www stryker com patents Do not stack more than 1 high ...

Страница 5: ...rn assist with the optional pendant control 17 Alarming call maintenance with the optional pendant control 19 Resetting CPR with the optional InTouch footboard 19 Starting and stopping max inflate with the optional InTouch footboard 19 Locking and unlocking with the optional InTouch footboard 19 Adjusting firmness with the optional InTouch footboard 20 Starting and stopping turn assist with the op...

Страница 6: ...2 2920 109 001 REV G www stryker com Table of Contents To obtain parts and service 35 Return authorization 35 Damaged product 35 International warranty clause 35 ...

Страница 7: ... frequently for proper positioning and to make sure that the support surface maintains proper inflation Always make sure that the tubing and wiring that is connected to the patient is long enough stabilized and secure during Lateral Rotation or Turn Assist Do not extubate or intubate patients during Lateral Rotation or Turn Assist The rotation functions could interfere with the performance of the ...

Страница 8: ...ents Failure to follow these cleaning instructions may void your warranty Always unplug the product power cord before cleaning or disinfecting to avoid the risk of shock CAUTION Improper usage of the product can cause injury to the patient or operator Operate the product only as described in this manual Do not modify the product or any components of the product Modifying the product can cause unpr...

Страница 9: ...tiple air cells filled and emptied by an integrated control unit to provide changes in the distribution of body weight for pressure relief PositionPRO offers turn assist See the specifications page for compatible frames Indications for use PositionPRO is intended to assist in the prevention and treatment of pressure injures including stages 1 2 3 4 unstageable injury and deep tissue injury that ar...

Страница 10: ...tationary 500 lb 226 8 kg Default 22 mmHg optimal pressure relief 200 lb 90 7 kg Turn Assist 300 lb 136 1 kg Power cord 15 foot 16 AWG cord with hospital grade plug for use with wall outlet 4 foot 16 AWG cord with hospital grade plug for use with accessory outlet Over current protection 2 fuses 5 x 20 mm 5 AMP Slo Blo 250 VAC 120VAC 50 60Hz 1AMP Two 250V 5A Fuses Output flow rate 12 5 LPM 0 4 SCFM...

Страница 11: ...ironmental conditions Operation Storage and transportation Temperature 104 F 40 C 50 F 10 C 158 F 70 C 40 F 40 C Contact information Contact Stryker Customer Service or Technical Support at 1 800 327 0770 Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA To view your operations or maintenance manual online see https techweb stryker com Have the serial number A of your Stryker product avail...

Страница 12: ...oduct illustration A B C D E F G H Figure 1 Product illustration A CPR straps E Handles B D rings F Head end C Foot box G Integration cable D Foot end H Power cord Date of manufacture The year of manufacture is the first four digits of the serial number ...

Страница 13: ...e onto the bed 2 Flip the foot section toward the head end 3 Turn the control box upside down and place the control box into the opening in the foot section of the support surface Figure 2 on page 9 Figure 2 Insert the foot box 4 Match the outer transparent tubes colors to the manifold colors and connect them Figure 3 on page 9 Green Blue Red White Black Yellow Figure 3 Connect tubes by color 5 Co...

Страница 14: ... the InTouch FL27 bed is QDF27 1534 Figure 5 Connect the pendant cable 7 Connect the power cord 4 in or 15 in and turn the switch to the ON or 1 position Note The switch is under the power cord 8 Fasten the straps over the power cord 9 Install the power cord in the two retaining clips Figure 6 on page 10 Figure 6 Retaining clips 10 Fasten the three retaining straps ...

Страница 15: ...box Figure 7 Installation complete and ready to flip 12 Fasten all four of the retaining straps to secure the support surface to the bed frame Installation on page 9Figure 9 on page 11 Figure 8 Strap locations Figure 9 Fasten the retaining straps 13 Install the clip for the pendant standalone optional control in one of the two locations listed below Figure 10 on page 11 and Figure 11 on page 12 a ...

Страница 16: ...spital grade three prong plug into a properly grounded receptacle to protect against shock hazard Grounding reliability can be achieved only when a hospital grade receptacle is used 14 Connect the power cord to the power source 15 Calibrate the bed following the procedures in the InTouch maintenance manual ...

Страница 17: ...sist and Lateral Rotation functions Positioning the patient WARNING Always center the patient on the support surface before starting functions Check the patient frequently for proper positioning and to make sure that the support surface maintains proper inflation Always make sure that the tubing and wiring that is connected to the patient is long enough stabilized and secure during Lateral Rotatio...

Страница 18: ...n a short period of time WARNING Always make sure that the operator has access to the CPR straps Pull the red CPR strap Figure 13 on page 14 that is located on the left and right side of the head end of the mattress You can activate one or either of the straps Note The CPR visual alarm LED turns red The CPR alarm LED continues to flash to indicate that a hazardous situation has been detected The C...

Страница 19: ...maximum level to aid in patient boosts and patient transfers To stop Max Inflate Press the Max Inflate button to stop inflation Note If Max Inflate has been on longer than 30 minutes the alarm will sound and the support surface will return to the default firmness setting Locking and unlocking the optional pendant control To Lock Press the Lock button to prevent options from being selected on the p...

Страница 20: ...nlock all pendant functions Adjusting firmness with the optional pendant control You can adjust the support surface firmness settings for patient comfort The product has five levels of firmness settings that range from 14 mmHg to 30 mmHg The default of 22 mmHg provides optimal pressure relief for patients up to 200 lb For average sized patients 18 22mmHg is recommended Notes For larger patients hi...

Страница 21: ...ise the bed siderails before starting Turn Assist to avoid the risk of patient fall Do not use dynamic mattress systems for stroke victims without a physician s order Do not turn a patient with unstable fractures unstable spinal cord injuries or those in skeletal traction Do not extubate or intubate patients during Lateral Rotation or Turn Assist The rotation functions could cause interference wit...

Страница 22: ...sitionPRO can turn a 300 lb patient up to 40 You can use the Turn Assist therapy with a timer so therapy can last up to 120 minutes without the need to calculate the time the patient has been receiving this therapy When patient is unattended the warning illuminates when you activate the Left or Right Turn Assist function and you have not raised the siderails Right and left locations are from the p...

Страница 23: ...larm Starting and stopping max inflate with the optional InTouch footboard To start Max Inflate Tap Max Inflate to inflate the air bladders to the maximum level to aid in patient boosts and patient transfers To stop Max Inflate Tap Max Inflate to stop the inflation Note If Max Inflate has been on longer than 30 minutes the alarm will sound and the support surface will return to the default firmnes...

Страница 24: ...o unlock all pendant functions Adjusting firmness with the optional InTouch footboard The support surface firmness settings may be adjusted for patient comfort requirements The product has five levels of firmness settings that range from 14 mmHg to 30 mmHg The default of 22 mmHg will provide optimal pressure relief for patients up to 200 lb For average sized patients 18 22mmHg is recommended Notes...

Страница 25: ...ims without a physician s order Do not turn a patient with unstable fractures unstable spinal cord injuries or those in skeletal traction Do not extubate or intubate patients during Lateral Rotation or Turn Assist The rotation functions could cause interference with the performance of the ancillary devices Do not leave the patient unattended during Turn Assist CAUTION Always secure all patient lin...

Страница 26: ...ist window C Stop Stop Turn Assist 5 Tap Increase or Decrease to increase or decrease the turn time 6 Tap Start button to start turn assist 7 Tap Back button to access the Stop button Notes PositionPRO can turn a 300 lb patient up to 40 Right and left locations are from the perspective of the patient If you do not tap Continue Turn Turn Assist will hold in a turned position for 2 hours after 2 hou...

Страница 27: ...ional InTouch footboard The alarm conditions Call Maintenance LED illuminates red and flashes to indicate a hazardous situation has been detected Review the Error Alarms section and call appropriate technical support personnel Transferring a patient to and from the support surface To transfer a Patient 1 Lock the brakes on both of the platforms that the patient is being transferred to and from bed...

Страница 28: ...able safety rules and institution protocols to for patient and operator safety 5 Plug the support surface power cord and the bed power cord into properly grounded hospital grade power outlets when you have reached the patient destination Note The support surface will maintain air pressure for up to four hours while unplugged Managing incontinence and drainage WARNING Always monitor the patient con...

Страница 29: ...ing linens or placing a patient on the surface Drying the product aids in preventing the performance of the product from being impaired Always follow hospital protocol for cleaning and disinfecting To clean the support surface covers between patient uses follow these steps in order 1 Unplug the support surface before cleaning 2 Lift up the foot section of the support surface to clean the bottom su...

Страница 30: ... stryker com Cleaning Cleaning the pendant Continued 4 Care must be taken to thoroughly rinse and dry the pendant following cleaning 5 Disinfect as necessary with a hospital grade disinfectant after cleaning has been completed ...

Страница 31: ...ent on the surface Failure to remove excess disinfectant may degrade the cover material Do not over expose the covers to higher concentration disinfectant solutions as these may degrade the covers Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the cover Prerequisite Minimum of two operators are required to disinfect the support surface Sugge...

Страница 32: ...unctions properly Left and Right CPR releases function properly All electrical connections function properly Power cord and plug are free of damage Air cells are free of excessive wear cracks tears or other damage Fire barrier cover for excessive wear All connectors are free of damage Notes It is recommended to replace cells every six months Replace worn or damaged components as necessary Product ...

Страница 33: ... Starting at the foot end roll the new fire barrier up and slide the fire barrier over the crib assembly Note Align the fire barrier on the crib before sliding over the crib assembly 9 Carefully slide the fire barrier up the crib assembly working from side to side to make sure that the fire barrier is tight on the crib assembly 10 Align the crib assembly on top of the bottom part of the cover Note...

Страница 34: ... Part Number Clip QDF2082 Pendant QDF2081 Pump 56 0541 Transformer 56 0047 Manifold Valve Assembly 56 0503 Control Board 56 0280 Top Cover Dartex Standalone Mattress 56 0274 Top Cover Dartex Integrated Mattress with FL27 InTouch Critical Care Bed 56 0528 Power Cord 4 foot QDF8087 Power Cord 15 foot QDF8088 Fan 56 0509 Sub Assembly Line Choke 56 0243 Sub Assembly Tilt Sensor 56 0287 Fire barrier 56...

Страница 35: ...000 4 11 5 Ut 95 dip in Ut for 0 5 cycle 40 Ut 60 dip in Ut for 5 cycles 70 Ut 30 dip in Ut for 25 cycles 5 Ut 95 dip in Ut for 5 sec 5 Ut 95 dip in Ut for 0 5 cycle 40 Ut 60 dip in Ut for 5 cycles 70 Ut 30 dip in Ut for 25 cycles 5 Ut 95 dip in Ut for 5 sec Main power quality should be that of a typical commercial and or hospital environment If the user of PositionPRO requires continued operation...

Страница 36: ...uation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer Note At 80 MHz and 800 MHz the separation distance for the higher frequency range applies Note These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures ob...

Страница 37: ...r frequency range applies Note These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people aField strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically ...

Страница 38: ... use in all establishments other than domestic and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic Emissions IEC 61000 3 2 Class A Voltage Fluctuations Flicker Emissions IEC 61000 3 3 Complies WARNING This equipment system is intended for use by health care professionals only This equipment system may cause radio i...

Страница 39: ...ded by Stryker In no event shall Stryker be liable for incidental or consequential damages To obtain parts and service Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives These representatives are factory trained available locally and carry a substantial spare parts inventory to minimize repair time Simply call your local representative or call...

Страница 40: ......

Страница 41: ......

Страница 42: ...2016 10 2920 109 001 REV G www stryker com Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA ...

Отзывы: