4
Warnings and Cautions
Please read this manual and follow its instructions carefully.
e words warning, caution, and note carry special
meanings and should be carefully reviewed:
Warning
Warnings indicate risks to the safety
of the patient or user. Failure to follow
warnings may result in injury to the
patient or user.
Caution
Cautions indicate risks to the
equipment. Failure to follow cautions
may result in product damage.
Note
Notes provide special information to
clarify instructions or present additional
useful information.
An exclamation mark within a triangle
is intended to alert the user of special
warnings or important operating and
maintenance instructions in the manual.
A lightning bolt within a triangle is
intended to warn of the presence of
hazardous voltage. Refer all service to
authorized personnel.
e incorrect use of any of the required tools and/or
techniques may risk damage to the equipment or injury to
the person carr ying out the procedure, subsequent operators,
or patient. Perform the repair ONLY if you have been
speci
fi
cally trained in the use of all pertinent equipment
and techniques. Stryker Endoscopy cannot continue to
guarantee compliance to UL, CSA, TUV, or other labeled
safety standards if service is performed by anyone other than
Stryker Endoscopy personnel.
To help avoid potential serious injury to the us er and the
patient and/or damage to this device:
1. Read this manual thoroughly, especially the
warnings, and be familiar with its c ontents prior to
using this equipment.
2. Never sterilize any part of the X8000 console.
3. Disconnect the X8000 from the electrical outlet
when inspecting the fuses.
4. Ensure that readjustments, modi
fi
cations, and/
or repairs are carried out exclusively by Stryker
Endoscopy or other authorized personnel.
5. Ensure that the electrical installation of the relevant
operating room complies with the applicable IEC,
CEC, and NEC requirements.
Federal law (United States of America) restricts this device
to use by, or on order of, a physician.
IMPORTANT SAFETY NOTICE
Before operating or performing any maintenance or
repairs on the X8000 Light Source, read the user guide
and service guide thoroughly and carefully. When using
any light source, such as the X8000,
fi
re and/or severe
injury may result to the patient, user or inanimate
objects, if the instructions in this manual are not
followed.
All light sources, including the X8000, can generate
signi
fi
cant amounts of heat at the scope tip, the scope
light post, the light cable tip, and/or near the light cable
adapter. Higher levels of brightness from the light source
result in higher levels of heat.
To avoid the risk of burns and/or
fi
re:
•
Always adjust the brightness level of the camera
and the monitor before adjusting the brightness
level of the light source.
•
Adjust the brightness level of the light source to
the minimum brightness necessary to adequately
illuminate the surgical site.
•
Adjust the internal shutter of the camera higher
in order to run the light source at a lower
intensity.
•
Avoid touching the scope tip or the light cable
tip to the patient, and never place them on top
of the patient, as doing so may result in burns to
the patient or user.
•
Never place the scope tip, the scope light post,
the light cable adapter, or the light cable tip on
the surgical drapes or other
fl
ammable material,
as doing so may result in
fi
re.
•
Always place the light source in Standby mode
whenever the scope is removed from the light
cable or the device is unattended.
e scope
tip, scope light post, light cable adapter, and
light cable tip will take several minutes to cool
o
ff
a
er being placed in Standby mode, and
therefore may still result in
fi
re or burns to the
patient, user, or inanimate objects.