Inspect the power cord receptacle for
bent pins or bent contacts.
Inspect the infrared communication
window for any dirt or debris.
Thoroughly remove any dirt or debris on
the infrared communication window. See
To Clean and Disinfect the Rover
50).
Six months
Inspect the replacement date on the label
of the fluid suction HEPA filter. The fluid
suction filter life is 500 hours.
Replace the fluid suction HEPA filter
every six months or as indicated on the
main control panel display. See
(page 12) and
N
No
otte
e -- If any component must be discarded, see
5
5..3
3.. T
To
o R
Re
ep
plla
acce
e tth
he
e H
HE
EP
PA
A F
Fiilltte
err
Failure to comply may cause infection and result in healthcare staff injury.
W
WA
AR
RN
NIIN
NG
G -- Do not touch the filter media. The media filters the air evacuated from fluid collection canister of the rover
before the air is vented. Failure to comply may cause media damage and result in patient and/or healthcare staff injury.
F
Fiig
gu
urre
e 4
40
0 –– P
Prre
evve
en
ntt D
Da
am
ma
ag
ge
e tto
o F
Fiilltte
err M
Me
ed
diia
a
EN
52
700000969828 Rev-AA
DRAFT
NOT
FOR
SALE