7. See
To Install Disposable Components
4
4..2
2.. T
To
o IIn
nsstta
allll D
Diissp
po
ossa
ab
blle
e C
Co
om
mp
po
on
ne
en
nttss
W
WA
AR
RN
NIIN
NG
G -- The manifold is for single patient use only. Do not reuse, process, service, modify, or repackage a single
use device.
•
The single-use device may not withstand chemical, chemical vapor, or high temperature sterilization processing.
•
Design features may make cleaning difficult.
•
Reuse may create a contamination risk and compromise structural integrity, resulting in operational failure during use.
Failure to comply may lead to infection or cross-infection and result in patient and/or healthcare staff injury.
W
WA
AR
RN
NIIN
NG
G -- Before use, inspect each disposable component for damage. Do not use any equipment if damage is
apparent.
W
WA
AR
RN
NIIN
NG
G -- Do not use a manifold that has been dropped or mishandled. Failure to comply may cause leakage and
result in patient and/or healthcare staff injury.
F
Fiig
gu
urre
e 2
20
0 –– D
Do
o N
No
ott D
Drro
op
p tth
he
e M
Ma
an
niiffo
olld
d
W
WA
AR
RN
NIIN
NG
G -- Do not remove the suction tubing from the manifold ports during use and at shutdown. Always leave the
suction tubing attached to the manifold ports after installation.
W
WA
AR
RN
NIIN
NG
G -- Do not use a Y-connector to add more than one suction line to each port. Failure to comply may cause
cross-contamination and result in patient injury.
F
Fiig
gu
urre
e 2
21
1 –– D
Do
o N
No
ott U
Usse
e a
a Y
Y--cco
on
nn
ne
ecctto
orr
N
No
otte
e -- If the collection of small specimens is required during the procedure, select a specimen collection manifold with
a collection tray. Obtain additional specimen collection trays for use, if necessary. See
N
No
otte
e -- If specimen collection is not required during the procedure, select the disposable 4-port manifold.
700000969828 Rev-AA
37
EN
DRAFT
NOT
FOR
SALE