![Stryker Medical Renaissance 1067 Скачать руководство пользователя страница 4](http://html.mh-extra.com/html/stryker-medical/renaissance-1067/renaissance-1067_maintenance-manual_1386143004.webp)
Preventative Maintenance
3
CLEANING
1.
Hand wash all surfaces of the bed with warm water and mild detergent. Dry thoroughly.
2.
Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow disinfectant to evaporate.
(Appropriate disinfectant for nylon Velcro should be determined by the hospital.)
NOTE
Quaternary Germicidal Disinfectants, used as directed, and/or Chlorine Bleach products, typically 5.25% So-
dium Hypochlorite in dilutions ranging between 1 part bleach to 100 parts water, and 2 parts bleach to 100
parts water are not considered mild detergents. These products are corrosive in nature and may cause dam-
age to your stretcher if used improperly. If these types of products are used to clean Stryker patient handling
equipment, measures must be taken to insure the stretchers are rinsed with clean water and thoroughly dried
following cleaning. Failure to properly rinse and dry the stretchers will leave a corrosive residue on the surface
of the stretcher, possibly causing premature corrosion of critical components. Failure to follow the above
directions when using these types of cleaners may void this product’s warranty.
BIANNUAL CHECKLIST
All fasteners and knobs secure (reference all assembly prints)
Siderails move and latch properly
All casters lock with brake pedal engaged (page 13)
Steer function working properly
All casters secure and swivel properly
Body restraints working properly
I.V. pole intact and operating properly (page 48)
Oxygen bottle holder intact and operating properly
Fowler operates and latches properly
Knee Gatch operates properly (page 39)
Trendelenberg/Reverse Trendelenberg operating properly
No rips or cracks in mattress cover
Ground chain intact
No leaks at hydraulic connections
Hydraulic jacks holding properly (page 8)
Hydraulic drop rate set properly (page 10)
Hydraulic oil level sufficient (page 9)
Lubricate where required, including the brake adjuster assembly and brake cam (page 13)
Serial No.______________
______________
______________
______________
______________
Completed By:_________________________________ Date:_____________
Содержание Renaissance 1067
Страница 15: ...1067 1 10 Side Control Base Assembly 14 FOOT END DETAIL A HEAD END See Detail A ...
Страница 17: ...1067 1 10 Side Control Base Assembly 16 FOOT END HEAD END ...
Страница 20: ...1067 2 10 End Control Base Assembly 19 See Detail A Detail A ...
Страница 22: ...1067 2 10 End Control Base Assembly 21 FOOT END HEAD END ...
Страница 62: ...DH 8 97 1067 1 16 REV E 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...