![Stryker Medical 1068 Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/stryker-medical/1068/1068_maintenance-manual_1386131005.webp)
Preventative Maintenance
4
CLEANING
1.
Hand wash all surfaces of the bed with warm water and mild detergent. Dry thoroughly.
2.
Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow disinfectant to evaporate.
(Appropriate disinfectant for nylon Velcro should be determined by the hospital.)
NOTE
Quaternary Germicidal Disinfectants, used as directed, and/or Chlorine Bleach products, typically 5.25% So-
dium Hypochlorite in dilutions ranging between 1 part bleach to 100 parts water, and 2 parts bleach to 100
parts water are not considered mild detergents. These products are corrosive in nature and may cause dam-
age to your stretcher if used improperly. If these types of products are used to clean Stryker patient handling
equipment, measures must be taken to insure the stretchers are rinsed with clean water and thoroughly dried
following cleaning. Failure to properly rinse and dry the stretchers will leave a corrosive residue on the surface
of the stretcher, possibly causing premature corrosion of critical components. Failure to follow the above
directions when using these types of cleaners may void this product’s warranty.
BIANNUAL CHECKLIST
All fasteners secure (reference all assembly prints)
Siderails move and latch properly (page 5)
Engage brake pedal and push on the stretcher to ensure all casters lock securely (page 15)
Steer function working properly
All casters secure and swiveling properly
Body restraints working properly
I.V. pole intact and operating properly (page 56–58.4)
Oxygen bottle holder intact and operating properly (page 32 or 33)
Fowler operating properly
Knee Gatch operating properly (page 48)
Trendelenberg/Reverse Trendelenberg operating properly
No rips or cracks in mattress cover
Arm boards intact and operating properly (page 5)
Arm board support levers intact and operating properly (page 40)
Ground chain intact
No leaks at hydraulic connections
Hydraulic jacks holding properly (page 10)
Hydraulic drop rate set properly (page 12)
Hydraulic oil level sufficient (page 11)
Lubricate where required, including the brake adjuster assembly and brake cam (page 15), and the
pin and block (”U” joint) on the Fowler assembly (page 44 & 45)
Serial No.______________
______________
______________
______________
______________
Completed By:_________________________________ Date:_____________
Содержание 1068
Страница 10: ...5 4 Notes ...
Страница 21: ...1068 1 10 Side Control Base Assembly 16 ...
Страница 23: ...1068 1 10 Side Control Base Assembly 18 ...
Страница 27: ...Notes 20 2 ...
Страница 28: ...1068 2 10 End Control Base Assembly 21 ...
Страница 30: ...1068 2 10 End Control Base Assembly 23 ...
Страница 42: ......
Страница 46: ...1068 30 10 Litter Assembly 34 HEAD END ...
Страница 58: ...1068 232 10 Enhanced Clearance Fowler Subass y 43 1 ...
Страница 60: ...1068 32 10 Crank Fowler Subassembly 44 ...
Страница 86: ...Notes 62 2 ...