Cleaning
5
Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow disinfectant to evaporate. (Ap-
propriate disinfectant for nylon Velcro should be determined by the hospital.)
In general, when used in those concentrations recommended by the manufacturer, either phenolic type or
quaternary type disinfectants can be used. Iodophor type disinfectants are not recommended for use be-
cause staining may result. The following products have been tested and have been found not to have a harm-
ful effect WHEN USED IN ACCORDANCE WITH MANUFACTURERS RECOMMENDED DILUTION.*
TRADE NAME
DISINFECTANT
TYPE
MANUFACTURER
*MANUFACTURER’S
RECOMMENDED
DILUTION
A33
Quaternary
Airwick (Professional Products Division)
2 ounces/gallon
A33 (dry)
Quaternary
Airwick (Professional Products Division)
1/2 ounce/gallon
Beaucoup
Phenolic
Huntington Laboratories
1 ounce/gallon
Blue Chip
Quaternary
S.C. Johnson
2 ounces/gallon
Elimstaph
Quaternary
Walter G. Legge
1 ounce/gallon
Franklin
Phenomysan
F2500
Phenolic
Purex Corporation
1 1/4 ounce/gallon
Franklin Sentinel
Quaternary
Purex Corporation
2 ounces/gallon
Galahad
Phenolic
Puritan Churchill Chemical Company
1 ounce/gallon
Hi–Tor
Quaternary
Huntington Laboratories
1/2 ounce/gallon
LPH
Phenolic
Vestal Laboratories
1/2 ounce/gallon
Matar
Phenolic
Huntington Laboratories
1/2 ounce/gallon
Omega
Quaternary
Airwick (Professional Products Division)
1/2 ounce/gallon
Quanto
Quaternary
Huntington Laboratories
1 ounce/gallon
Sanikleen
Quaternary
West Chemical Products
2 ounces/ gallon
Sanimaster II
Quaternary
Service Master
1 ounce/gallon
Vesphene
Phenolic
Vestal Laboratories
1 1/4 ounce/ gallon
Quaternary Germicidal Disinfectants, used as directed, and/or Chlorine Bleach products, typically 5.25% So-
dium Hypochlorite in dilutions ranging between 1 part bleach to 100 parts water, and 2 parts bleach
to 100 parts water are not considered mild detergents. These products are corrosive in nature and
may cause damage to your stretcher if used improperly. If these types of products are used to clean
Stryker patient handling equipment, measures must be taken to insure the stretchers are rinsed with clean
water and thoroughly dried following cleaning. Failure to properly rinse and dry the stretchers will leave a cor-
rosive residue on the surface of the stretcher, possibly causing premature corrosion of critical components.
NOTE
Failure to follow the above directions when using these types of cleaners may void this product’s warranty.
REMOVAL OF IODINE COMPOUNDS
This solution may be used to remove iodine stains from mattress cover and foam footrest pad surfaces.
1. Use a solution of 1–2 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area.
Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to
soak or stand on the surface.
2. Rinse surfaces which have been exposed to the solution in clear water before returning bed to service.
Содержание 1020
Страница 19: ...18 715 1 250 Side Control Base Assembly with Brakes ...
Страница 30: ...Notes 29 ...
Страница 33: ...Side Control Base Assembly with Jacks 32 Assembly part number 715 1 260 ...
Страница 37: ...Side Control Base Assembly with Jacks 36 Assembly part number 1210 2 10 ...
Страница 40: ...Notes 39 ...
Страница 41: ...End Control Base Assembly with Jacks 40 FOOT END HEAD END Assembly part number 716 1 270 ...
Страница 50: ...49 1020 1 10 Litter Assembly ...
Страница 51: ...50 Litter Assembly ...
Страница 65: ...64 1020 22 10 X Ray Tray Assembly Right Side Loading ...
Страница 67: ...66 1020 21 10 X Ray Tray Assembly Left Side Loading AF BL BR BP ...
Страница 93: ...DH 11 97 1020 32 34 REV K 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...