7.4.2
Changing the cooling fluid
Note
The cooling unit fluid contains additive and grinding residue and you must not
dispose of it into the waste water drain.
Cooling fluid must be disposed of in compliance with local safety regulations.
Emptying the recirculation tank
1.
Remove the recirculation tank from the compartment under the machine.
2.
Disconnect the water inlet hose from the pump and connect the extra piece of
hose (supplied).
3.
Place the other end of the hose in a container of a suitable size.
4.
From the
Maintenance
menu select
Empty recirculation tank
.
5.
Press
F1
to start the pump.
6.
The pump stops automatically. To stop it manually, press
Stop
7.
Follow the on-screen instructions.
CAUTION
Avoid skin contact with the cooling fluid additive.
7.5
Annually
Inspecting the dosing arm
1.
Lift the dosing arm and move it to the edge of the machine.
2.
Inspect the nozzles.
3.
If needed, clean the nozzles.
Inspecting the safety cover
Hint
If the machine is used for more than one 7-hour shift per day, carry out inspection
more often.
1.
Visually inspect the safety cover for signs of wear or damage such as cracks, dents, or
damage to the sealing edge.
7 Maintenance and service - AbraPol-30
AbraPol-30
43