Objet260 User Guide
6 Operating and Maintaining the Objet260 3D Printer
DOC-37004 Rev. E
6-66
6. Connect the cable from the sensor to the UV meter.
7. Turn on the UV meter, and wait until “0 0 0.0” appears on the display.
8. In the wizard screen, confirm that the sensor is positioned correctly, and
click
Next
.
Continue with "UV Measurement " below.
UV Measurement
The UV lamps power up and stabilize. (This takes several minutes.) Then, the
print block passes over the sensor and the wizard compares the measured
radiation to the recommended level for each UV lamp and printing mode.
During this process, which takes 20–40 minutes, you need to monitor the
progress as displayed in the wizard screen.
Figure 6-98 UV measurement progress
When using a stand-alone radiation meter, reset the meter and enter the
reading (measurement) when you are prompted to do so.
Figure 6-99 UV Measurement dialog box