![Storz OR1 FUSION WO300 Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/storz/or1-fusion-wo300/or1-fusion-wo300_instructions-for-use-manual_1382924091.webp)
Electromagnetic compatibility
Instructions for use • KARL STORZ OR1 FUSION
®
, Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR
91
Test fre-
quency
MHz
Frequency
band
MHz
Radio service
Modulation
Immunity
test level
V/m
Compliance
level
V/m
1,970
LTE band 1, 3, 4,
25,
UMTS
2,450
2,400 –
2,570
Bluetooth,
WLAN 802.11 b/g/
n,
RFID 2450,
LTE band 7
Pulse modulation
217 Hz
28
28
5,240
5,100 –
5,800
WLAN 802.11 a/n
Pulse modulation
217 Hz
9
9
5,500
5,785
9.5 Table 3 – Test levels for radiated and conducted
immunity tests
Guidelines and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user
of the product should make sure that it is used in such an environment.
Interference
immunity tests
EN/IEC 60601
test level
Compliance
level
Electromagnetic environment – guide-
lines
Conducted HF
disturbances
acc. to IEC
61000-4-6
3 V
rms
150 kHz to
80 MHz
3 V
rms
Portable and mobile HF communications
equipment should be used no closer to
any part of the product, including cables,
than the recommended separation dis-
tance calculated from the equation appli-
cable to the frequency of the transmitter.
Recommended separation distances:
d = 1.2 √P
Where P is the rated power of the trans-
mitter in watts [W] according to the infor-
mation provided by the transmitter manu-
facturer and d is the recommended sepa-
ration distance in meters [m].
Field strengths from fixed HF transmitters
as determined by an electromagnetic site
survey
a
should be less than the compli-
ance level in each frequency range
b
.
d = 1.2 √P
80 MHz to 800 MHz
d = 2.3 √P
800 MHz to 2.5 GHz
Interferences may occur in the vicinity of
equipment marked with the following sym-
bol:
Radiated HF dis-
turbances acc.
to IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m