Electromagnetic compatibility
Instructions for use • C-MAC
®
Pocket Monitor • DRE117_EN_V1.0_10-2020_IFU_CE-MDR
33
10.3 Table 2 – Test levels for proximity fields from HF
wireless communications equipment
Test fre-
quency
MHz
Frequency
band
MHz
Radio service Modulation
Immunity
test level
V/m
Compliance
level
V/m
385
380 – 390
TETRA 400
Pulse modula-
tion 18 Hz
27
27
450
430 – 470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz sine
wave
28
28
710
704 – 787
LTE band 13 &
17
Pulse modula-
tion
217 Hz
9
9
745
780
810
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE band 5
Pulse modula-
tion
18 Hz
28
28
870
930
1,720
1,700 – 1,990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE band 1, 3,
4, 25,
UMTS
Pulse modula-
tion
217 Hz
28
28
1,845
1,970
2,450
2,400 – 2,570
Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE band 7
Pulse modula-
tion
217 Hz
28
28
5,240
5,100 – 5,800
WLAN 802.11
a/n
Pulse modula-
tion
217 Hz
9
9
5,500
5,785
10.4 Table 3 – Test levels for radiated and conducted
immunity tests
Guidelines and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user
of the product should make sure that it is used in such an environment.
Interference immunity
tests
EN/IEC 60601 test
level
Compliance
level
Electromagnetic envi-
ronment – guidelines
Conducted HF distur-
bances acc. to IEC
61000-4-6
3 V
rms
150 kHz to 80 MHz
3 V
rms
Portable and mobile HF
communications equip-
ment should be used no
closer to any part of the