8 Installation
8.1 Putting into service
Before operating this device for the first time:
•
Inspect the device visually (
chapter 13.1 Recurrent Tests)
.
•
Check the mains voltage (compare the data on the type label with the availa-
ble mains voltage). An incorrect mains voltage can lead to destruction of the
device
National regulations may require other tests before this device can be put into
service. If required, additional tests for electrical safety should be carried out ac-
cording to
chapter 13.1 Recurrent Tests.
8.2 Installation of the Control unit
For safe installation, the control unit is equipped with a universal attachment de-
vice. With this, the device can be securely attached to infusion stands as well as
to medical standard rails.
CAUTION
Risk of injury!
When installing the ASTOPAD DUO310 control unit on an infusion stand, ob-
serve the instructions from the infusion stand manufacturer regarding maximum
load and tilting stability.
8.2.1 Attachment to infusion stands
Fig. 4 Attachment to infusion stands
1. Turn the hand wheel counterclockwise to open the
attachment device.
2. Turn the hand wheel clockwise to lock the attach-
ment device to the infusion stand.
3. Check that the control unit is firmly fixed.
8.2.2 Attachment to medical rails
Abb. 5 Attachment to medical rails
1. Hang the ASTOPAD control unit obliquely from
above with the attachment device into the standard
rail.
2. Fix the ASTOPAD control unit by tightening the
hand wheel to the standard rail.
3. Check that the control unit is firmly fixed.