WARNING
Risk of radio interference!
•
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is neces-
sary, this equipment and the other equipment should be observed to verify
that they are operating normally.
•
Use of accessories, other than those specified or provided by the manufact-
urer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper
operation.
•
Portable RF communications equipment (including peripherals such as an-
tenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of ASTOPAD, including cables specified by the manu-
facturer. Otherwise, degradation of the performance of this equipment could
result.
3.3 Cautions
CAUTION
Risk of injury!
•
Federal law (USA) restricts this device to sale by or on the order of a Licensed
Healthcare Practitioner.
•
When installing the ASTOPAD DUO310 control unit on an infusion stand, ob-
serve the instructions from the infusion stand manufacturer regarding maxi-
mum load and tilting stability.
•
When ASTOPAD is used on the OR table, the OR table must be prepared
according to the customary regulations and guidelines.
•
Never insert pointed or sharp objects into the applied parts or damage the
surface of the parts in any other way.
•
Damage to the Applied part can cause overheating, therefore:
- Do not use bleach solution with hypochlorite for disinfection of the applied parts.
- Other than the cleaning and disinfection procedures described in this IFU shall
not be applied without the manufacturer’s authorization.
•
Do not continue to use ASTOPAD applied parts beyond the labeled Use by Date.
CAUTION
Risk of hypothermia!
•
If the alarm shut-off of the ASTOPAD is triggered at an output, the entire system
will switch off.
•
If thermally conductive materials, such as water, gel, and similar substances, are
used and were not pre-heated, the patient’s body temperature may reduce as a
result once the ASTOPAD applied parts are switched off.