Sterling Sterilisers

Acuatic Vacuum 12 B

User Instructions

Содержание Acuatic Vacuum 12 B

Страница 1: ...Sterling Sterilisers Acuatic Vacuum 12 B User Instructions...

Страница 2: ...n EC No 1882 2003 0086 indicates that the Notified Body employed to check conformation with the above is BSI This symbol indicates that the number next to it is the serial number of the steriliser Thi...

Страница 3: ...Logger Daily Maintenance 4 DailyTesting 4 Weekly Testing 4 Servicing 4 Annual Inspection 4 Annual Testing 4 Getting Started GettingStarted 5 Filling 5 Water Quality 5 Draining 5 Options 5 Operation S...

Страница 4: ...onstruction structural parts are manufactured from stainless steel or aluminium Impossible to overheat innovative engineering eliminates downtime due to overheating Easy water fill just pour into the...

Страница 5: ...ising Drying 138 C 134 C 1 0 2 Pressure bar 1 Time Air removal and heating Drying Sterilising B type A door seal B chamber with tray rack and trays C water filling spout D touch screen display E clean...

Страница 6: ...d instruments the internal surfaces of the steriliser and the steriliser door may be hot Do not touch these parts until cooling has taken place and exercise care when handling just sterilised instrume...

Страница 7: ...is observed Weekly Testing Run the automatic air leakage test This test will show that the steriliser is capable of pulling and holding a certain level of vacuum The display will indicate a pass or f...

Страница 8: ...this is not recommended because of the risk of the heaters becoming scaled Draining When the used water reservoir is full the display will ask for it to be drained This will need to be done before the...

Страница 9: ...If the N type cycle is enabled you will be offered a choice of starting N type or B type cycles Processing When a cycle is running the display will show the cycle time elapsed the operating temperatu...

Страница 10: ...ep the Add shortcut to desktop checkbox ticked To import cycle data from the SD memory card to your computer run the Securilog application by double clicking the Securilog icon on your desktop Remove...

Страница 11: ...with the Medical Devices Directive 4 A Product Registration and Customer Survey form Please take the time to print this out complete it and return it to Sterling Sterilisers so that we can keep our re...

Страница 12: ...apped pouched or not Limits Design pressure 1 to 3 bar Design temperature 2 to 144 C Water Water used per cycle 200 to 700 ml Clean water reservoir volume 2 litres Used water reservoir volume 2 litres...