3

High Power Vacuum Pump

 is supplied to produce cycle 

vacuum pulses that remove air and moisture from chamber. 
Direct drive rotary vane pump is quiet (<60 dB) with low 
vibration. Powerful 3 HP (2.2 kW) motor produces a 
displacement of 53 CFM (90 m

3

/hr) and helps alleviate moisture 

sensitivity in biodecontamination unit. Biodecontamination unit 
operating pressure is from atmospheric pressure down to less 
than 1 Torr.

CONTROL DESCRIPTION

The VHP MD140X Biodecontamination Unit is equipped with 
an Allen-Bradley Compact Logix

 (PanelView Plus

 6 1000 

Display

1

) control. PLC features a 10.4" color TFT display with 

touch screen.

Screens are color coded for operator convenience as follows:

• Control Screens - 

» Condition Phase - Green
» Biodecontamination Phase - Blue
» Aeration Phase - Violet

• Service Screens - Light Blue

• Option Screens - Purple

• Alarm Screens - Red 

NOTE: This Biodecontamination Unit permits no manual 
control of the Biodecontamination Cycles.

1

 CompactLogix™ and PanelView™ Plus 6 1000 are trademarks of 
Allen-Bradley, a Rockwell Automation Company.

CYCLE DESCRIPTION

The  VHP MD140X Biodecontamination Unit is equipped with 
three programmed Biodecontamination Cycles: Lumen Cycle 
(55 minutes), Non Lumen Cycle (28 minutes) and Flexible 
Cycle (35 minutes). Each Biodecontamination Cycle proceeds 
through three phases: CONDITION, BIODECONTAMINATION 
and AERATION.

Basic description of a Biodecontamination Cycle (example):

CONDITION

 — This cycle phase is a set time vacuum 

pulse to remove air and moisture from the chamber. When 
setpoint is reached, load is tested for acceptable moisture 
content. If content is acceptable, cycle proceeds. If not, 
Condition pulse repeated.

NOTE: If Condition phase fails the second moisture check, 
the cycle Aborts. 

BIODECONTAMINATION

 — This cycle phase is a series of 

four pulses. Each pulse consists of: vacuum pulled to 
setpoint; Vaprox 59 Hydrogen Peroxide Sterilant vapor 
drawn into chamber; hold for programmed time; filtered air 
is introduced to setpoint; hold for programmed time; deep 
vacuum pulled to setpoint.

AERATION

 — This cycle phase pulls a vacuum to 

setpoint and continues to evacuate for programmed time 
to reduce chamber vapor concentration. Once Aeration 
phase is complete, chamber pressure returns to 
atmospheric and chamber door is unlocked. 

NOTE: Automatic door locking mechanism keeps chamber 
locked during entire Biodecontamination Cycle.

BIODECONTAMINATION ASSURANCE 
PRODUCTS

Vaprox 59 Hydrogen Peroxide Sterilant

 – 59% stabilized 

aqueous solution of hydrogen peroxide designed for use with 
VHP Biodecontamination Units and Accessories (EPA Reg. No. 
1043-123). Order PB033US (950 mL).

Steraffirm

®

  NB305 Process Indicator

 – Chemical indicator 

designed for use with hydrogen peroxide vapor.

Steraffirm

®

 PCC051 Process Indicator

 – Chemical indicator 

designed for use with hydrogen peroxide vapor.

Spordex

®

 NA330 Biological Indicator

 – E6 

Geobacillus 

stearothermophilus

 inoculated stainless-steel coupon 

designed for use with hydrogen peroxide vapor.

Spordex

®

 NA114 Biological Indicator Media

 – Culture media 

designed for use with Spordex and Spordi

®

 biological 

indicators.

ACCESSORIES

Seismic Tie Down Kit 

– Required for seismic installations for 

safe operation of Biodecontamination Unit.

Trays And Organizer Clips

 – Trays are lightweight, durable 

and available in a variety of sizes to fit your instruments while 
organizer clips anchor instruments in-place for transportation.

Tray Mats 

– Mats are lightweight, durable and feature flexible 

prongs which loosely grip instruments.

Pouches

 – Vis-U-All™ pouches combine durable Tyvek with 

a clear 2.0 mL film to offer a wide variety of pouch sizes and 
styles in heat-seal, self-seal and customizable roll tubing.

PREVENTIVE MAINTENANCE

Customers are encouraged to contact STERIS Life Sciences 
concerning annual maintenance programs. Preventive 
maintenance, adjustments and replacement of worn parts are 
provided on a scheduled basis to help ensure optimal 
equipment performance and help minimize untimely and costly 
interruptions. STERIS Life Sciences maintains a worldwide staff 
of well-equipped, factory-trained technicians to provide these 
services, as well as on-site installation, training and expert 
repair services. Contact STERIS Life Sciences

2

 for details.

2

(800) 444-9009 or www.sterislifesciences.com

CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH 
APPLICABLE LOCAL AND NATIONAL CODES AND 
REGULATIONS.

Содержание VHP MD140X

Страница 1: ... compatible with a broad range of materials and has rapid Biodecontamination Cycle times There are no toxic by products created by the Biodecontamination Cycle only water vapor and oxygen are produced The VHP MD140X Biodecontamination Unit performs three programmed Biodecontamination Cycles Lumen Cycle approximately 55 minutes to complete Non Lumen Cycle approximately 28 minutes to complete Flexib...

Страница 2: ...er right corner Each bar represents approximately four cycles Unit is available with single door in either a freestanding or recessed configuration Unit installation requires no plumbing ventilation or air supply only a dedicated electrical connection is needed A power cord is supplied for this connection Unit is equipped with a racking system for articles to be biodecontaminated Unit is equipped ...

Страница 3: ...nt and continues to evacuate for programmed time to reduce chamber vapor concentration Once Aeration phase is complete chamber pressure returns to atmospheric and chamber door is unlocked NOTE Automatic door locking mechanism keeps chamber locked during entire Biodecontamination Cycle BIODECONTAMINATION ASSURANCE PRODUCTS Vaprox 59 Hydrogen Peroxide Sterilant 59 stabilized aqueous solution of hydr...

Страница 4: ...t Floor surface must be hard and level 4 Biodecontamination Unit should not be installed in an area not compatible with oxidizers Consult the MSDS regarding hydrogen peroxide sterilant 5 STERIS recommends maintaining and operating Biodecontamination Unit in area where temperature ranges from 41 104 F 5 40 C and has ventilation system exchanging area air at least 10 times per hour 6 STERIS recommen...

Страница 5: ...5 ...

Страница 6: ...mber Equipment Drawing Title 387358 924 Single Door Recessed 387358 923 Single Door Cabinet SINGLE DOOR RECESSED MODEL SHOWN Drawing is not to scale Dimensions are typical Dimensions are in inches mm Not For Installation SIDE VIEW PLAN VIEW FRONT VIEW 30 1 2 775 DOOR SWING ...

Страница 7: ...umber Equipment Drawing Title 387358 924 Single Door Recessed 387358 923 Single Door Cabinet SINGLE DOOR CABINET MODEL SHOWN Dimensions are in inches mm Not For Installation SIDE VIEW PLAN VIEW FRONT VIEW Drawing is not to scale Dimensions are typical 30 1 2 775 DOOR SWING ...

Страница 8: ...440 354 2600 800 444 9009 www STERISLifeSciences com SD980 2013 STERIS Corporation All rights reserved 03 01 13 This document is intended for the exclusive use of STERIS Customers including architects or designers Reproduction in whole or in part by any party other than a Customer is prohibited ...

Отзывы: