INDICATIONS FOR USE

Document # M4070EN.2016-03, Rev. B

©2016 STERIS Corporation.
All rights reserved.

The V-PRO 60 Low Temperature Sterilization System, with VAPROX

®

 HC Sterilant, is a vaporized hydrogen peroxide 

sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal 
medical devices used in Healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen 
Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical 
devices sensitive to heat and moisture.

a The validation studies for all lumen configurations were conducted using a maximum of twelve (12) lumens per load. Hospital loads should not exceed the maximum number 
of lumens validated by this testing. The validation studies were performed using a validation load consisting of one instrument tray with instrument organizers and mat and two 
pouches for a total weight of 11 lbs (5.0 kg). 

b  The validation studies were performed using a validation load consisting of one instrument tray with instrument organizers and mat and one pouch for a total weight of 12 lbs  
(5.4 kg).

c  The validation studies were conducted with one flexible endoscope, packaged into a tray with silicone mat, instrument organizers and light cord (if not integral to scope) and  
no additional load.

V-PRO 60

Sterilizer Cycle

Indications for Use

Total Load

Lumen Cycle

Can sterilize

a

:

• Instruments with diffusion-restricted spaces such as the

hinged portion of forceps and scissors

• Non-lumened devices including non-lumened rigid and

semi-rigid endoscopes

• Medical devices, including single, dual and triple

channeled rigid and semi-rigid endoscopes, with the
following configurations

a

:

1. Single or dual lumen devices with stainless steel

lumens that are ≥ 0.77 mm (~1/32”) internal diameter
(ID) and ≤ 410 mm (16-9/64”) in length

2. Triple lumen devices with stainless steel lumens that are:

≥ 1.2 mm (~3/64”) ID and ≤ 275 mm (~10-55/64”) in length
≥ 1.8 mm (~5/64”) ID and ≤ 310 mm (~12-13/64”) in length

or

      ≥ 2.8 mm (~7/64”) ID and ≤ 317 mm (12-31/64”) in length

11 lbs (5.0 kgs) 

and 

12 Lumens

Non Lumen Cycle

Can sterilize

b

:

• Non-lumened devices including non-lumened rigid, semi- 
      rigid and flexible endoscopes and non-lumened devices
      with stainless steel or titanium diffusion-restricted spaces 
      such as the hinged portion of forceps and scissors.

12 lbs (5.4 kgs)

Flexible Cycle

Can sterilize

c

:

• One flexible surgical endoscope or bronchoscope with

a light cord (if not integral to endoscope) and mat without
any additional load. The flexible endoscope may be a:

Single or dual lumen device with lumens that are > 1 mm
(~3/64”) ID and < 990 mm (38-63/64”) in length

NA

1

Teflon is a registered trademark of the E.I. du Pont de Nemours and Company.

2

da Vinci

®

 is a registered trademark of Intuitive Surgical, Inc. 