6-4
122994-513
Operator Manual
Routine Maintenance
6.3 Daily Cleaning
1.
After last cycle of day, allow unit to cool. Then, remove and clean
debris screens in bottom of wash chamber and suction filter at
bottom of sump. Always clean screens while they are still wet,
before foreign matter dries.
2.
Remove riser valve and inspect for debris. Brush and rinse
under tap water if necessary.
6.4 Weekly Cleaning
On a weekly basis, perform the following routine cleaning proce-
dures:
1.
Clean washer exterior as follows: Using a general cleaner for
general stains, a stainless-steel stain remover for stubborn stains
and a stainless-steel polish to keep equipment looking like new.
a. Using a damp cloth or sponge, apply cleaner in a back-and-
forth motion, rubbing in same direction as surface grain.
b. Thoroughly wipe off cleaner.
c. Polish surface with a clean, dry, lint-free cloth.
2.
Wash chamber interior with a moderately alkaline detergent solu-
tion. Rinse with tap water and dry with a lint-free cloth.
NOTE: If interior is stained, use a general cleaner to remove
general stains, or a stainless-steel stain remover for stubborn
stains. Using a damp cloth or sponge, apply cleaner in a back-and-
forth motion, rubbing in same direction as surface grain. Thoroughly
wipe off cleaner and polish with a clean, dry, lint-free cloth.
3.
Clean wash chamber rotary spray arm assemblies as follows:
a. Remove rotary spray holder from top of wash chamber.
NOTE: Before removing rotary spray holder, note the following:
1) There are two loose bushings, one under and one over rotary
spray hub. When lowering rotary spray assembly, bushings
fall. Ensure to hold on to bushings while removing rotary
spray assembly.
2) Do not remove spray arms from rotary spray arm hub.
b. Rinse each spray arm assembly under running water to
clean out sediment.
c. Use a fine wire (approximately wire gauge of a paper clip) to
clean sediment from spray nozzles. Rinse under running water.
d. Place bushings on rotary spray arm hub and reattach spray
arm assembly to top of wash chamber.
4.
Clean rotary spray arms on accessories in same way as cham-
ber spray arms.
WARNING–SLIPPING HAZARD:
To prevent slips, keep floors
dry. Promptly clean up any
spills or drippage. For spills
or drippage of detergents or
other chemicals, follow safety
precautions and handling pro-
cedures set forth on deter-
gent or chemical label and/or
Material Safety Data Sheet
(MSDS).
WARNING–SLIPPING HAZARD:
To prevent slips, keep floors
dry. Promptly clean up any
spills or drippage. For spills
or drippage of detergents or
other chemicals, follow safety
precautions and handling pro-
cedures set forth on deter-
gent or chemical label and/or
Material Safety Data Sheet
(MSDS).
CAUTION – POSSIBLE EQUIP-
MENT DAMAGE: Use nonabra-
sive cleaners when cleaning
unit. Follow directions on con-
tainers and rub in a back-and-
forth motion (in same direction
as surface grain). Abrasive
cleaners will damage stain-
less steel. Cleaners rubbed in
a circular motion or applied
with a wire brush or steel wool
on door and chamber assem-
blies can be harmful to stain-
less steel. Do not use these
cleaners on painted surfaces.
Содержание RELIANCE 400
Страница 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Страница 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Страница 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Страница 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Страница 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Страница 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Страница 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Страница 105: ......
Страница 106: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 JIT 4 USE 800 548 4873 ...