![Steris RELIANCE 400 Скачать руководство пользователя страница 81](http://html.mh-extra.com/html/steris/reliance-400/reliance-400_operators-manual_1374370081.webp)
5-29
Cycle and Control Value Programming
Operator Manual
122994-513
5.9.2 Verifying Chemical
Concentration
See Figure 5-9.
To verify or establish chemical concentration, operator may collect
water from sample port located behind lower service panel under
sump.
1.
On Cycle menu, press
SELECT CYCLE
touch pad until desired
cycle where injection time was modified appears:
__ indicates flashing position
2.
Press
CYCLE/START
touch pad to run cycle.
3.
While cycle is running, let water recirculate and when Wash
treatment starts, press
STOP/RESET
touch pad.
4.
To collect a sample, place a small container under sample port
and turn knob counterclockwise. To close sample port, turn knob
clockwise.
5.
Analyze chemical concentration from sample. Adjust concentra-
tion if needed, following procedure described in
.
LIGHT MEDIUM
HEAVY
CYCLE 4
SELECT
CYCLE
CYCLE
START
STOP
RESET
REF.: # 920-012-111
Figure 5-9. Sample Port
Sample Port
Lower Service Access Panel
Sump
Содержание RELIANCE 400
Страница 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Страница 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Страница 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Страница 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Страница 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Страница 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Страница 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Страница 105: ......
Страница 106: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 JIT 4 USE 800 548 4873 ...