1-1
Safety Precautions
Operator Manual
920011-649
1
The following
Safety Precautions
must
be observed when operating or servicing this equipment. WARNING
indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For
emphasis, certain
Safety Precautions
are repeated throughout the manual.
It is important to review ALL Safety
Precautions before operating or servicing the
Reliance™ EPS Processor
.
Strictly following these Safety Precautions enhances the Customer’s ability to safely and effectively utilize the unit
and helps avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop their own
safety policies and procedures to enhance and complement these Safety Precautions.
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
WARNING – ELECTRIC SHOCK AND/OR BURN HAZARD:
Bag-in-a-Box compartment door can support a maximum of 4.5 kg (10 lb). Do not place heavy objects
on Bag-in-a-Box compartment door.
In an emergency, stop cycle by pressing STOP touch pad twice to abort cycle. Wait for processor to
complete Abort sequence. Wear appropriate Personal Protective Equipment (PPE) when reaching into
chamber.
Lack of appropriate training could result in injury or equipment damage.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment.
Contact STERIS to schedule preventive maintenance or obtain the necessary Maintenance Manual if
preventive maintenance is done by the Customer.
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service
personnel. Nonroutine maintenance performed by unqualified personnel or installation of unauthorized
parts could cause personal injury, result in improper equipment performance or result in costly damage.
Contact STERIS regarding service options.
The performance of the 510k cleared Reliance EPS Endoscope Processing System is validated as a
system including components defined by STERIS in the Operator Manual and Maintenance Manual for
the Reliance EPS Endoscope Processing System. Substitution of unauthorized components can
potentially lead to personal injury, damage or premature failure of the product and result in a unit
configuration that is inconsistent with the validated, FDA cleared product. Altering the equipment or
using unapproved component’s may void STERIS’ warranty. DO NOT USE components that are not
validated as part of the Reliance EPS Endoscope Processing System.
When working with compressed air, flying debris and particles may present a hazard to personnel in
immediate area. Personal eye protection must be worn when working with compressed air.
Disconnect all utilities to processor before servicing. Do not service processor unless all utilities have
been properly locked out. Always follow local electrical codes and safety-related work practices.
SAFETY PRECAUTIONS
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...