7-8
764333-268
b.
Remove table column cover per
.
c.
Remove appropriate base cover per
d.
Stack shims (P150832-657) under indicated
caster wheel until fit is tight. Note the quantity
used.
e.
Carefully tilt and secure table enough to allow
removal or caster needing shims.
f.
Remove flat washer, lockwasher and nut
fastening caster to table base (see Figure 7-20).
g.
Position shims (quantity determined in
Step 7d
) between caster and table base.
h.
Reassemble in reverse order. Torque nut to
20 N•m (177 lbf•in).
7.11 FLOOR LOCK MANIFOLD
ASSEMBLY REPLACEMENT
Replace a faulty floor lock manifold assembly as
follows:
1. Release floor locks. Ensure table is on casters and
not floor locks.
2. Turn Main Power Switch to OFF and remove
battery fuse (refer to supplied Operator Manual for
more information).
3. Remove table column cover per
.
4. Remove appropriate base cover per
.
5. Carefully lift battery #2 from table base and place
on top of battery #1.
6. Remove all hydraulic and electrical connections
from manifold assembly. Note and record various
connections for ease of reassembly. Discard all
hydraulic screws and sealing washers.
NOTE: Refer to Cmax Surgical Table Hydraulics
figure in the Illustrated Parts Breakdown section for
hose identifications.
7. Remove both stainless-steel socket head cap screws
fastening floor lock manifold to manifold spacer
(see Figure 7-21).
8. Replace manifold assembly.
9. Clean manifold area and reassemble in reverse
order. Replace sealing washers and torque two-way
hydraulic screws to 12 N•m (106 lbf•in).
NOTE: Replace (do not reuse) hydraulic screws and
sealing washers (see Figure 7-21). Ensure bolt
threads, manifold threads and hose connectors are
free of rubber seal particles.
Figure 7-21. Replace Floor Lock Manifold
7.12 HYDRAULIC PUMP REPLACEMENT
Replace a faulty hydraulic pump as follows:
1. Position tabletop to level, slide at zero.
2. Raise tabletop to full height.
3. Turn Main Power Switch to OFF and remove
battery fuse (refer to supplied Operator Manual for
more information).
4. Remove table column cover per
5. Remove appropriate base cover per
6. Remove stainless-steel socket head cap screws and
washers fastening rear base cover support bracket
to table base. Set fender washers, screws and bracket
aside (see Figure 7-22).
7. Disconnect electrical harness from battery charger
(connector TB3).
8. Remove and discard hydraulic screws and sealing
washers from top of pump.
9. Carefully lift pump assembly from table base.
10. Repair or replace pump and reinstall by reversing
procedure. Replace sealing washers and torque
hydraulic screws to 12 N•m (106 lbf•in).
NOTE: Replace (do not reuse) hydraulic screws and
sealing washers (see Figure 7-22). Ensure bolt
threads, manifold threads and hose connectors are
free of rubber seal particles.
11. Operate table through several movements to
ensure proper operation.
12. Check hydraulic oil level when table is lowered
completely and tabletop level. Level should be
65 mm (2-9/16").
Fastening sc
Hydraulic Screw
Mounting Screw
Two-Way Hydraulic
Screw and Sealing
Washers
Electrical
Connection
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......