Cycle and Control Value Programming
26. Press
VALUE arrows
(up or down) to toggle between YES and NO.
Selecting YES generates a printout of cycle phase values once
SAVE
VALUES
touch pad or
CHANGE VALUES
touch pad is pressed.
27. At any point during change values sequence, operator has the option to
either continue in Change Values mode or to save values and exit Change
Values mode.
• To exit Change Values mode, press
SAVE VALUES
touch pad. Control
saves changed values and returns to selected cycle menu. Display
shows:
CYCLE 9
CYCLE 10
CYCLE 11
CYCLE 12
__ Indicates flashing position.
• To continue in Change Values mode, press
CHANGE VALUES
touch
pad. Change operating values option appears. Display shows:
CHANGE OPERATING
VALUES?
NO
__Indicates flashing position.
NOTE: Refer to Section 5.2.2, Programming Operating Values, for
details on continuing in the Change Values mode.
CHANGE
VALUES
SAVE