4-11
Operating Instructions
Operator Manual
122998-741
6.
Water-Alkaline neutralizer solution is mixed:
pH TOO HIGH
MIXING WATER
7.
Sump water is pumped to drain. If temperature is guaranteed,
actual sump temperature alternates with guaranteed set temper-
ature.
CYCLE 10
(G=)128.5 F
RINSE 1 DRAIN
1:00
4.3.11 Rinse 2 Phase
1.
Hot water from building supply fills sump. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature.
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time will only count
down when solution temperature is equal to or greater than set
point.
CYCLE 10
(G=)137.4 F
RINSE 2 FILL
2:00
2.
Rinse 2 solution is recirculated through oscillating jet system for
programmed amount of time. If temperature is guaranteed,
actual sump temperature alternates with guaranteed set temper-
ature.
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time will only count
down when solution temperature is equal to or greater than set
point.
CYCLE 10
(G=)140.0 F
RINSE 2
5:00
3.
Printer records time and solution temperature in chamber sump
at beginning and end of recirculation.
RINSE 2
10:21:55A
141.5
10:26:55A
140.2
**************************************
WARNING – BURN HAZARD:
Water discharge may be
extremely hot.