![Steris AMSCO 400 Series Скачать руководство пользователя страница 106](http://html.mh-extra.com/html/steris/amsco-400-series/amsco-400-series_operators-manual_1374315106.webp)
7-12
P129394-020
Operator Manual
Cycle and Control Value Programming
7.5 Change Machine
Setup
All changes are made to displayed settings using touch screens. No
mechanical adjustments to the sterilizer are necessary.
Generally, the Setup options are used to change the way the
sterilizer operates. The control has an Access Code Security feature.
If the Access Code is enabled, all or some of these options can be
secured or “locked out” by the supervisor (see Access Code later in
this section for more information on Access Code). All Setup options
are accessed from the Setup Menu.
A summary of the setup options that can be adjusted are listed in
. Each value is detailed in this section.
Table 7-2. Change Machine Setup
CHANGE MACHINE SETUP
Once the access code has been set, all cycles values and Change Machine Setup options
can be selected to lockout, making them unavailable for change by an operator without
entering the Access Code.
This setting is used to secure any of the Change Values options under the access code.
Once a function is locked out, the Access Code must be entered before the setting can be
accessed or changed.
This setting permits the operator to program the sterilizer to automatically shut off its steam
and water at the end of the work day, to conserve utilities. Shut down and power-up times
can be programmed for any time of the day on Weekdays or the Weekend.
NOTE: Sterilizers equipped with an integral steam generator automatically run a generator
flush-and-drain cycle one hour before sterilizer’s scheduled restart time.
If the sterilizer is equipped with an integral steam generator, use Generator Flush Control to
set up two options: FLUSH TIME AND FLUSH DURATION.
This option can be used to select one of three factory set languages as the default for displays
and printouts. Available languages are ENGLISH, SPANISH and FRENCH. A BILINGUAL
options allows for easy changes between English and Spanish, or English and French.
This is used to assign a six-character, alphanumeric code to the sterilizer. This code appears
in the heading of all printouts.
This setting allows the control to display times using either AM/PM format or 24 hour format.
This setting allows the sterilizer to print a condensed version of the cycle printout to conserve
on paper usage.
This option allows adjustment of ALARM, END OF CYCLE and TOUCHPAD signals tones.
Tones can be independently adjusted to one of three volume levels. ALARM signal tone
cannot be turned off.
This is used to select the temperature and pressure units displayed and printed out by the
sterilizer. The sterilizer is capable of displaying temperature as either Fahrenheit or Celsius;
pressure can be displayed as psig/inHg, millibar or psia.
The sterilizer can be programmed to change the way the date is displayed and printed. The
normal setting is to display Month/Day/Year (M/D/Y); but this can be changed to
Year/Month/Day (Y/M/D), or Day/Month/Year (D/M/Y).
Sterilizer can be set to automatically furnish a duplicate printout of each cycle at the end of
the cycle. First line will always read — DUPLICATE PRINT — and complete printout of cycle
data will be furnished.
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...