“plus” Systems incorporating BA-ICE-CM controller:
1. UV intensity (%):
The Cobalt “Plus” series of products incorporate a UV sensor which detects the
discrete 254 nm wavelength of the UV lamp. This information is relayed to the Cobalt
“Plus” controller and is the default display shown in “% UV output”. The system
will display the UV output between 50 to 99 percent. When the system drops below
50%, a low UV warning is displayed as and alternately flashes (at 2 second
intervals) back to the actual UV level. Eg. Additionally, the system will
supply an intermittent audible tone (2 seconds on, 2 seconds off), during low UV
conditions.
Note:
UV levels of ...
to
Indicates the system is functioning within normal a normal
operating range.
to
Indicates the UV level is still within a safe level, however cleaning
or lamp replacement may soon be required.
to
Indicates the UV level is nearing the point of unsafe UV intensity,
UV system should be immediately serviced.
<
Indicates the UV level has now reached a level that is unsafe. At
this level the water should not be consumed. The system/water
supply should be examined to determine the reason for the low
UV level of the UV intensity. At this level, the solenoid output has
been activated and if a solenoid is installed, water will cease to
flow.
DEFErrAL
– To temporarily defer the audible alarm during a low UV alarm, press the
push-button “RESET” switch and hold for five seconds. This will mute the audible
alarm condition for 12 hours.
This advanced warning system has been installed to provide
you with the optimum protection against microbiological
contamination in your water.
Do NoT DISREGARD THE WARNING SIGNALS.
The best way to ensure optimum UV performance is to have
the water microbiologically tested by a recognized testing
agency on a regular basis.
12