2. Safety 

Instructions 

 
 
The folowing safety instructions will be 
repeated at all relevant sections of the 
Operating Instructions and must be complied 
with at all times. 
 

Indicates a danger that, in the event 
of non-compliance, could lead to 
life-threatening injuries for the 
patient and/or the operator. 
 

Indicates a caution that, in the event 
of non-compliance, could lead to 
malfunctioning of, damage to, or 
defects on the equipment, and 
possibly endanger the patient. 
 

 

2.1. Danger 

warnings 

 

The 

MedAir

 may be operated only 

by trained and authorized personnel. 
The 

MedAir

 must be operated in 

accordance with these Operating 
Instructions.  
 
The 

MedAir

 may not be operated in 

areas exposed to a danger of 
explosion. 

 

The equipment may not be operated 
without a micro-filter. 
 
 
The air intake filter on the rear side 
may not be covered by any objects. 
 
 
In order to ensure an optimal supply 
of fresh air and air circulation the 
equipment is only to be operated in 
an adequate sized room which is 
well-ventilated and preferably dust-
free. Otherwise this can be seriously 
damaging to the equipment. 

 
 
 
 
 
 
 
 
 
 
 
 

 
 

2.2. Caution 

warnings 

 
For medical/technical equipment 
with electrical connection, DIN EN 
60601-1 and VDE 0751-1 are to be 
complied with. Accordingly, such 
equipment may only be put into 
initial operation by the manufacturer 
or authorized technical support 
personnel. 

 
 

Содержание 161061000

Страница 1: ...EPHAN GmbH Medizintechnik D 56412 Gackenbach Kirchstr 19 Tel 06439 9125 0 Fax 06439 9125 111 info stephan gmbh com www stephan gmbh com MedAir Mobile Compressed Air Unit h Operating and service Instructions ...

Страница 2: ...mbly of the system Should there arise any problems with the functioning of the system please get in touch with the Technical Support team at F Stephan GmbH or the authorized dealer from whom you have received the equipment and received training in its use The manufacturer guarantees the safety and reliability of the MedAirs only when the equipment is operated in compliance with the Operating Instr...

Страница 3: ...al terminology 4 1 7 Technical Data 5 2 Safety Instructions 6 2 1 Danger warnings 6 2 2 Caution warnings 6 3 Description of the Design and Functioning 7 3 1 Front view 7 3 2 Side view left 8 3 3 Rear view 8 4 Preparing the Equipment for Operation 9 4 1 Setting up the MedAir 9 4 2 Connection to the mains power supply 9 5 Operation 10 6 Problems Causes and Remedies 11 7 Care and Servicing 12 7 1 Cle...

Страница 4: ... cardboard cartoning can be reused or disposed of as waste paper The immediate wrapping around the system is of CFC free material It can be disposed of along with the foil using the recyclable waste bin 1 5 Introduction The German Equipment Regulations MPG the Medical Products Operations Act MPBetreibV and the law governing technical materials stipulate that the attention of the user operator must...

Страница 5: ...moisture and humidity store dust free and free from frost Electrical supply Mains power 230V 50Hz Power consumption 250 W Current consumtion approx 5A Power output 600 VA Fuse protection 3 15 AT Noise level 45 dB A Operating pressure 3 5 bar Output flow 0 45 l min General Protection rating I Inspection maintenance annually or every 3500 operating hours Dimensions L x W x H 500x510x1150 Weight 50 k...

Страница 6: ...orized personnel The MedAir must be operated in accordance with these Operating Instructions The MedAir may not be operated in areas exposed to a danger of explosion The equipment may not be operated without a micro filter The air intake filter on the rear side may not be covered by any objects In order to ensure an optimal supply of fresh air and air circulation the equipment is only to be operat...

Страница 7: ...ter bacterial filter and an additional condensation water extractor The MedAir is mobile vibration and sound insulated A side rail 25x10 mm allows the mounting of additional equipment Monitoring and operation are via a control panel with mains power switch operating hours counter fuses and pressure gauge 3 1 Front view 1 Power switch ON OFF 2 Fuses 3 15 AT 3 Operating hours counter 4 Pressure gaug...

Страница 8: ...1 Air intake filter matting 2 Bacterial micro filter with the intake sound damper positioned behind 3 Moisture micro filter with moisture micro filter insert 4 Mains power connection cable 5 Drainage valve Depending on fill level the condensation water is to be drained from catch tank For doing this the drainage valve lever 5 must be pushed upwards 1 1 2 3 4 5 ...

Страница 9: ...ty exceed 70 it is recommended that the equipment only be operated in air conditioned areas The air intake filter positioned at the rear of the unit may not be covered by any objects The rear side of the unit must be at least 10 cm from the wall 4 2 Connection to the mains power supply Check to determine whether the mains power voltage matches the power voltage as indicated on the rating plate on ...

Страница 10: ... MedAirs is activated which ensures adequate cooling of the unit s interior Generation of compressed air begins approximately 10 seconds after that The compressor of the MedAirs builds up an operating pressure of 3 8 bar in approximately 4 seconds The pressure gauge on the front side of the unit shows the pressure at the AIR outlet The outlet connection coupling is positioned on the left side of t...

Страница 11: ... filter is dirty or clogged Leakage in line Defective compressor Gas usage is too high Replace micro filter Contact Technical Support Reduce gas usage Pressure continuously below 3 5 bar Intake sound damper is dirty Defective pressure safety valve Replace intake sound damper Replace defective presurre safety valve Not functioning Mains power failure The equipment automatically restarts after a res...

Страница 12: ... replaced when found dirty or at least every 6 months 7 2 Safety technical inspections Safety technical inspections as well as checking the quality of air generated are to be carried out annually or after 3500 operating hours by the manufacturer or by the authorized Technical Support team of F STEPHAN 7 3 Servicing For reasons of safety it is recommended that the servicing of the MedAirs be carrie...

Страница 13: ...ng 4 47 x 1 78 for nippel coupling AIR when leakage detected Items 1 to 6 can be ordered in a set Art No 1 610 61 500 which is sufficient for three years The set additionally includes a replacement parts case and diverse small parts Optional accessories Item Article number Article description Quantity 1 1 602 61 112 Connection hose for compressed air length 3 m 1 2 1 600 40 010 Shock proof socket ...

Страница 14: ...s a warranty of 24 months beginning at purchase date Modifications or repair work may only be carried out by the manufacturer F Stephan GmbH or authorized technicians Otherwise the warranty becomes void Warranty shall also become void as a result of improper use and treatment of the equipment ...

Страница 15: ...Notes Operating Instructions MedAir February 2007 Personal Notes ...

Отзывы: