Menu bar position----Choosing the
menu’s position on the screen(on the left or right of the screen)
(
3
)
Menu bar hide icon----click this icon, the menu bar will be hidden in
the corner of the screen, and click the icon again to display the menu bar.
1.3
Admin login
Click the System setting icon
in the Assist Tools bar, click the drop-down
button and choose the Amin, then click login. Every time starts the microscope,
user must finish the Admin login first. Or user can’t do the calibration. We have
done the calibration in advance before shipping the microscopes to customers.
1.4
Calibration
1.4.1 New Calibration
The user can select the “calibration method” as needed, click the drop-down button, select
“line calibration” or “circle calibration” to enter the creation calibration state, for example, select
the three-point circle calibration method and place the circular calibration plate under the lens.
After the image is adjusted clearly, select three points on the outer edge of the circle to draw a
circle, check how well the circle is drawn and the circle of the calibration board, if you are not
satisfied, you can redraw the circle until you are satisfied. Then enter the magnification of the
current lens and the actual size of the calibration circle in the dialog box. At this time, the current
calibration information will appear in the menu "Calibration" dialog box. After calibration, if the
lens is a zoom lens, you can switch to another magnification to continue the calibration. Repeat
the operation on the software to perform another magnification calibration. Repeat the above
calibration in sequence to complete the calibration between different magnifications of the lens.