6 | Espire Elbow Technical Manual - Classic Plus, Classic and Basic
Espire Elbow Technical Manual - Classic Plus, Classic and Basic | 7
SECTION 2 -
INPUTS (CLASSIC
PLUS)
2.1 Input Overview
The Espire Elbow is compatible with many types of inputs,
offering versatile control strategies.
Example List of Supported Inputs:
• A/C Remote Electrodes*
• D/C Cased Electrodes
• Linear Transducers
• Touch Pads
• Switches (Single-State, Dual-State, Bump, etc)
• Other Items Not Listed (Contact Steeper Group)
*Not compatible with Steeper A/C Electrodes with TruSignal
TM
Note: For an input to be compatible, the shape of the
plug connector must match the receptacle on the input
board.
2.2 Input Connector Board
There are 4 plug receptacles on the input board. The table
below lists the location, the type of input that can be used, and
its setup type.
Input Connector Board - Top View
Board Location
Input Option
1
Battery
2
Electrode-Close
3
Electrode-Open
4
Switch
2.3 Connecting Cables to the Input Board
1. Apply silicone grease to the plug connectors before
inserting into the board.
2. When attaching cables, note the proper orientation. The
connectors are “keyed” or asymmetrical to assure proper
alignment. The connector should plug in easily and is
held in place with friction.
3. Once the cables are attached, apply more silicone grease
on top of the connectors to prevent moisture from
entering receptacles.
4. When removing cables, pull close to the connector to
avoid pulling on the wires. Wires that become loose
could cause intermittent operation.
1
2
3
4