Страница 1: ...Espire Elbow Classic Plus Classic and Basic Technical Manual ...

Страница 2: ...AMINATION COLLAR AND CLAMP RING 6 1 Orientation of the Lamination Collar and Clamp Ring 6 2 Internal External Rotation 6 3 Attaching the Lamination Collar to the Elbow 6 4 Determining Proper Orientation on Test Socket 6 5 Final Adjustment with the User SECTION 7 MANUAL LOCK 7 1 Manual Lock Overview 7 2 Operating Manual Lock SECTION 8 CABLE MOUNTING 8 1 Cable Mounting Overview 8 2 Setting Up the Ca...

Страница 3: ...des provisions for myoelectric control of terminal devices Espire Classic Mechanical elbow with manual lock and forearm counterbalance for lift compensation Body powered only Espire Basic Mechanical elbow with manual lock and forearm spring assist for lift compensation Body powered only Features 1 3 Pre Installation Checklist What s in the box Hardware Espire Elbow Lamination Collar and Clamp Ring...

Страница 4: ...nector Board There are 4 plug receptacles on the input board The table below lists the location the type of input that can be used and its setup type Input Connector Board Top View Board Location Input Option 1 Battery 2 Electrode Close 3 Electrode Open 4 Switch 2 3 Connecting Cables to the Input Board 1 Apply silicone grease to the plug connectors before inserting into the board 2 When attaching ...

Страница 5: ...n 4 2 Below Elbow Battery Location The Espire Classic Plus provides a compartment on the bottom side of the elbow for housing the 3500S batteries and if needed other manufacturers batteries Other manufacturers batteries can be installed in this compartment providing the battery dimensions fit without modifying the arm 1 Route the battery wire through the hole inside of the compartment into the for...

Страница 6: ...273 mm 10 73 in 5 Minimum Length w rotator Elbow Centre 222 mm 8 75 in 225 mm 8 86 in 6 Minimum Length w out rotator Elbow Centre 200 mm 7 88 in 225 mm 8 86 in 7 Maximum Cut Area w rotator Distal End 25 mm 1 00 in 48 mm 1 875 in 8 Maximum Cut Area w out rotator Distal End 48 mm 1 875 in 48 mm 1 875 in 4 2 Protective Foam Insert Classic Plus A foam insert is installed prior to shipping to protect t...

Страница 7: ...itch Note If mounting the battery in the forearm please refer to section 3 for further details 1 2 2 5 3 Quick Disconnect Wrists Classic Plus Wiring The hand open hand close and power output cables will plug into a single coaxial unit which fits through the centre of the quick disconnect wrist 1 Plug the orange hand open cable into the left plug labeled 2 then plug the green hand close cable into ...

Страница 8: ...rotation This feature is to prevent input wires from being twisted and potentially damaged The anti rotation stop pin allows for 100 external 100 internal rotation for a total range of 200 Info Normal human range of motion is 30 externally and 135 internally Humeral Friction Fasteners 6 3 Attaching the Lamination Collar to the Elbow 1 Disassemble the clamp ring by removing the fasteners 2 Place th...

Страница 9: ... Manual Lock Overview The Espire features a mechanical locking mechanism that can be locked or unlocked while under load with a maximum support of 25lbs 11kg The manual lock cable is attached to the upper elbow assembly near the midline The lock can be engaged disengaged every 10 13 positions A strain relief is included to reduce the likeliness of damage due to excessive pull on the lock cable The...

Страница 10: ...ing point is the first hole closest to elbow centre The closer to the joint the bracket is located the more force is required to flex the elbow 8 2 Setting Up the Cable Mounting Drill Mounting Hole A 17 64in or 6 75mm drill bit and collar are provided to control the depth of the hole that is drilled Carefully drill the mounting hole Do not drill beyond the forearm shell surface or contact the inte...

Страница 11: ...hat if the adjustment dial is at maximum tension and the arm is raised to a horizontal level the elbow could suddenly flex Caution To prevent injury users should ensure the elbow is in the maximum flexed position when donning or doffing Adjustment Dial Example left arm with medial adjustment location 9 2 Counterbalance Adjustment Direction Adjustment Result Turn the dial posteriorly to increase th...

Страница 12: ...shumeral shoulder disarticulation and forequarter Adequate muscle activity and range of motion of shoulder joint to adequately control a body powered prosthetic device 11 3 Safety Please ensure that the user is fully aware of all the safety instructions before they leave the clinic Note Possible technical damage Info Basic information regarding this product Caution Possible risk of accident or inj...

Страница 13: ... Consequences of product deterioration Wear and tear on system components can lead to malfunction of the Espire Elbow resulting in a risk of injury The service life of this device is 2 years for device parts and accessories Caution Water and Humidity The electrical and mechanical systems of your Espire Elbow are not water resistant Water must be prevented from entering the Espire Elbow Be careful ...

Страница 14: ...on Source The requirements for accreditation and market surveillance relating to the marketing of products Medical Device Directive 765 2008 EC 768 2008 EC MDD 93 42 EEC Articles 4 11 12 17 Annex II Temperature limit ISO 15223 1 Reference no 5 3 7 Storage humidity range ISO 15223 1 Reference no 5 3 8 Protection against solid foreign objects of 12 5 mm diameter and greater and protection against ve...

Страница 15: ...zzo Drive Leeds UK LS10 1DF Steeper Group Unit 3 Stourton Link Intermezzo Drive Leeds LS10 1DF Tel 44 0 870 240 4133 Email customerservices steepergroup com SteeperUSA 8666 Huebner Road Suite 112 San Antonio TX 78240 Tel 1 210 481 4126 Email inquiries steeperusa com STPPR126 Issue 1 November 2020 ...