Содержание ELEC50

Страница 1: ...Electrode Technical Information ELEC50 ELEC60 ...

Страница 2: ...nufacturers Proportional output signals Suitable for child or adult systems High sensitivity 2000 100 000 fold and range 90 450Hz Able to function with threshold control using signals as low as 10V Interference protection from common power sources depending on electrode type and high frequency emitting devices An Electrode Gain Control EGC is included to allow control of the myoelectric signal pro...

Страница 3: ...F position For external batteries disconnect battery from housing Standard compliance The CE marking has been carried out in accordance with the requirements of the 93 42 EEC guidelines 8747 Steeper operate a UKAS approved quality management system and fully complies with the requirements of BS EN ISO 9001 2008 ISO27001 2005 ISO50001 TickIT Issue 5 This certifies that Steeper meets the appropriate...

Страница 4: ...signal and biocompatibility ELEC50 ELEC60 Provides interference protection in countries operating 60Hz mains power systems see map Appropriate control device for prosthetic devices from all major manufacturers Titanium contacts for optimum conductivity of signal and biocompatibility Light grey Dark grey Mains power system map 50Hz electrode ELEC50 should be used in these areas 60Hz electrode ELEC6...

Страница 5: ...rature range 5 o F to 140 o F 15 o C to 60 o C Max humidity 95 Maximum atmospheric pressure 104Pa Power supply 5V to 16V Bandwidth 90 450Hz A E25590 Lamination kit E25591 Vacuum kit Steeper batteries recommended 1 3 1 Vacuum kit 2 Lamination kit 3 Electrode ELEC50 or ELEC60 4 Adjustment tool 5 Spare suspension legs 6 User guide card 7 Technical information 4 5 2 ...

Страница 6: ... of the following parts can be ordered using the appropriate part number A guide for using the lamination kit and vacuum forming kit is available at www steepergroup com or www steeperusa com A printed copy is also available upon request Cable Retaining Plug B30819 Suspension Legs B31032 Lamination Kit Vacuum Kit Cable Retaining Plug Gain Level Adjustment Tool ELECSP Do not force the dial beyond t...

Страница 7: ...lectrode 1 2 3 4 5 6 Use NLU0010 silicone grease to seal the cable retaining block Remove any excess after connection Ensure blue side of electrode cable ECA 300 600 or 1000 is facing towards the electrode WARNING ...

Страница 8: ...e two replacements included Firstly remove them as shown and then replace as shown 1 2 3 2 3 1 Apply a small amount of NAD0107 glue to the underside of the electrode suspension leg in the area shown Electrode suspension leg reattachment Remove any excess adhesive after insertion Insert the suspension leg into the cavity ...

Страница 9: ...40 4133 Returns Returns If you need to return your product for any reason please contact your local distributor or customerservices steepergroup com or inquiries steeperusa com providing full details A returns number will be issued allowing your items to be dealt with promptly Disposal If you no longer require your product please return it to your local supplier or Steeper SteeperUSA with a reques...

Страница 10: ...Notes ...

Страница 11: ...it 3 Stourton Link Intermezzo Drive Leeds LS10 1DF 44 0 870 240 4133 customerservices steepergroup com www steepergroup com SteeperUSA 3619 Paesanos Parkway Suite 200 San Antonio TX 78231 USA 210 481 4126 inquiries steeperusa com www steeperusa com ...