Troubleshooting and Frequently Asked Questions 

What are the major problems encountered when setting up a system? 

 

1..If in doubt call Static Solutions and we will run through the system while you are on the keyboard. 
2..Installing the footplate cable so it protrudes from the edge of the stand will cause problems because the plug will 

bend and become dislodged from the jack. 

3.Low battery if not using the AC power supply. 

 
 

 

How do I hook up the two pig tail wires which are on the “special” meter to open doors? 

The relay circuit is normally open. When the test results are passed, the two leads will close because of the 

activation of the low voltage relay in the meter and allow the passage of a maximum of 24 volts at 

500milliamps through the relay to the door. See the connection diagram listed above. 

 

Why does the 9 volt battery get extremely hot ? 

 

Do not use a alkaline battery when using the power supply. Us a rechargeable battery if the power supply is 

used. 
 

 

What can be done if the readings for the foot wear are low? 

Check the footplate wire. Insert the wire completely into the meter. Workers wearing leather shoes that are 
wet will produce low readings. 

DO NOT CLEAN OFF THE FOOTPLATE WITH TOPICAL ANTISTAT 

CLEANERS. 

 

 

 

More troubleshooting tips are available through Static Solutions’ customer service Meter 
Department. Special systems can be ordered from Static Solutions, which will open doors, or have 

an upper limit of 1000 megohms.

 

 
 
 

Limited Warranty: 

 
Static Solutions warrants for a period of one year from the date of purchase the CT-8900 will be free of 

defect in material. Within the warranty period the meters will be replaced free of charge. Any meters 

returned shall be shipped prepaid to Static Solutions along with a return authorization number and 

proof of purchase. 
 

Exclusions. 

 

The above warranty will not apply to defects or damage due to accidents, misuse, design alterations, 

neglect, operator error, failure to clean or maintain the meters. 
 
Limitations: 

 

In no event shall Static Solutions be responsible for any loss, damage, injury, direct or consequential, 

arising out of the use of or the inability to use the meters. Before using, users shall determine the 
suitability of the product for their intended use. Users will assume all risk and liability what so ever in 

connection therewith. 

 

 

Static Solutions CT-8700 Combo Tester Instructions 

10 

4/22/2008 

 

Содержание OHM-STAT CT-8700

Страница 1: ...and plate are optional For OHM STAT CT 8700 Wrist and Foot Strap Combo Tester Kit SERIAL NUMBER COMPANY PRICE 25 00 US Email contactus staticsolutions com www staticsolutions com Static Solutions Inc 331 Boston Post Road East Marlboro Massachusetts 01752 Tel 508 480 0700 FAX 508 485 3353 Static Solutions CT 8700 Combo Tester Instructions 1 6 22 2008 ...

Страница 2: ...TION 1 DOOR RELAY WIRING AND SPECIFICATIONS 1 TESTER ADJUSTMENTS 1 DIP SWITCH SETTINGS 1 OPERATOR INSTRUCTIONS 1 SPECIFICATIONS 3 Troubleshooting and Frequently Asked Questions 3 Limited Warranty 3 Exclusions 3 Limitations 3 Congratulations on purchasing your CT 8700 wrist and foot strap testing kit This is the finest tester available on the market Static Solutions CT 8700 Combo Tester Instruction...

Страница 3: ...s Display Red yellow and green LED buzzer Accuracy 10 Weight Approx 1 0 lb 454 grams Meter alone with battery Battery type Life Typically over 20 000 measurements with a 9 volt alkaline battery Continuous power may be run with an external 9 15 VDC power supply Remove alkaline battery before running with an external power supply CE listed NIST certificate included Environment 320F to 1000F 00C to 3...

Страница 4: ...scharge Wrist straps and heel grounders simply act as a wire to drain off charges However to prevent accidental electrocution of the wearer should the wearer inadvertently touch an external high voltage source a high resistance is built into the strap The resistance is typically between 5 to 2 megohms Thus straps and heel grounders have to be tested for two different malfunction conditions If ther...

Страница 5: ...ons technical staff or refer to www StaticSolutions com for technical support literature Technical questions should be directed to contactus StaticSolutions com Fasten the foot stand to the footplate Place the foot stand on top of the footplate Insert the small Allen head bolts through the holes in the circular plate on the bottom of the stand Position these bolts into the T nuts located in the bo...

Страница 6: ...turns green the test was also a pass Make sure the toggle switch is in the correct testing position If either the red Too High or red Too Low LED illuminates and the buzzer activates or the computer screen turns red the worker must adjust and re test the wrist strap cord immediately To test a faulty cord leave the cord plugged into the meter and detach the wrist cord from the strap Press the wrist...

Страница 7: ...USTMENTS The CT 8900 can be adjusted to various electrical resistance specifications by opening the meter case and pressing the dipswitches corresponding to the desired test ranges see chart below The resistance ranges depend on what test standard i e EOS CECC the user follows The user must know what resistance range values are acceptable for their floors shoes and heel toe grounders and according...

Страница 8: ... off off 10M off ON off 25M ON ON off 35M off off ON IEC CECC default 50M ON off ON 75M off ON ON 100M ON ON ON US default Wrist High sw6 sw7 sw8 2M off off off 5M ON off off 10M off ON off US default 25M ON ON off 35M off off ON IEC CECC default 50M ON off ON 75M off ON ON 100M ON ON ON Figure 1 Dip switch see table for settings A special 1000 megohm unit and door opening mechanism is also availa...

Страница 9: ...rist Low 0 none ON default 75M sw3___sw4 __sw5 Foot High 20M off off off ON off off 50M off ON off ON ON off 100M 250M 350M 500M 750M 1000M off off ON ON off ON off ON ON ON ON ON default Wrist High 20M sw6___sw7 __sw8 off off off default 50M ON off off 100M 250M 350M 500M 750M 1000M off ON off ON ON off off off ON ON off ON off ON ON ON ON ON Static Solutions CT 8700 Combo Tester Instructions 9 4...

Страница 10: ...the meter Workers wearing leather shoes that are wet will produce low readings DO NOT CLEAN OFF THE FOOTPLATE WITH TOPICAL ANTISTAT CLEANERS More troubleshooting tips are available through Static Solutions customer service Meter Department Special systems can be ordered from Static Solutions which will open doors or have an upper limit of 1000 megohms Limited Warranty Static Solutions warrants for...