6 PKG08861-UM733.1
GENERAL INFORMATION
Basic Operating Instructions
1. Turn on your tool, check the upper left of the display to see
if the battery symbol is showing. If the battery symbol is
showing (Fig.1A) then go to the “Rechargeable Battery Care and
Maintenance” section, on pg.9. If there is no battery symbol
visible then the battery is charged.
2. Lightly clean the contact point.
3. Fasten the Micrometer into the appropriate holding device if
applicable.
4. You can turn the Micrometer on by either pressing one of the
buttons or moving the spindle.
5. If applicable, pick the unit of measure, inch or millimeter by
pressing the IN/mm button. Note: standard metric micrometers do
not have this function available.
6. Close the Micrometer by moving the spindle and the anvil
measuring faces together and then zero the tool with a quick
press on the Zero / ABS Button.
7. Zero Sequence: When you press the ZERO button the display
will show a dashed line as seen in (Fig.1B). The dashed line will
incrementally disappear from left to right (Fig.1C); Make sure not
to move the spindle during this time. This is a visual reminder to
wait for the tool to zero out. This will happen each time the tool is
zeroed, and takes less than a second to complete.
8. Carefully place the piece to be measured between the spindle
and the anvil measuring faces. Move the spindle to the piece
until contact is made, continue until the friction thimble no
longer moves the spindle. The measurement will be shown on
the display.
9. The Micrometer will automatically enter sleep mode in 30
minutes, if left unattended.
Button Placement and Functions
The functions printed in yellow; LIMITS and PRESET, are used in conjunction with
the SHIFT/SET button also printed in yellow. To enable these functions press the
SHIFT/SET button first. The SET (S) icon will appear in the top left corner of the
display, and then press the button for the required function. To Power on the tool
move the spindle or press any button. The tool will turn off automatically.
1
Wireless
Button
Turns on the Wireless module allowing communication
with other Wireless devices.
2
USB Mini
B Port
This is the USB port to charge the tool’s battery.
3
SHIFT/SET
Dual function button used to enable the, Preset and the
Limits function. When enabled the SET (S) icon will be
displayed on the top left corner of the display.
4
PRESET
Press the SHIFT/SET button and then press the PRESET
button to enable the PRESET function.
5
ZERO
Press and release the ZERO button and the display will
zero. The spindle must not be moved until the Zero
Sequence has elapsed.
6
ABS
Press and hold the button for 2 seconds enables ABS
mode. Press and hold for 2 more seconds to exit the
ABS mode.
7
IN/mm
Toggles the display between English or Metric units.
8
LIMITS
Press the SHIFT/SET button then press the LIMITS button
to enable the function.
9
Hold
Press and hold the button for a second will freeze the
display and the word “HOLD” will appear. A second press
and the display will update to the current position reading.
10
Send
A quick press and release of the button will send a reading
from the tool when the wireless is enabled.
Содержание SW733.1BXFLZ
Страница 1: ...Trust Is In The Name User Manual PKG08861 UMW733 1 ...
Страница 3: ...PKG08861 UM733 1 3 W733 1 ELECTRONIC MICROMETERS ...
Страница 9: ...PKG08861 UM733 1 9 FUNCTIONS ...
Страница 18: ...18 PKG08861 UM733 1 SPECIFICATIONS AND ACCESSORIES ...
Страница 24: ...24 PKG08861 UM733 1 ESPAÑOL ...
Страница 38: ...38 PKG08861 UM733 1 FRANÇAIS ...
Страница 52: ...52 PKG08861 UM733 1 PORTUGUÊS ...
Страница 66: ...66 PKG08861 UM733 1 中文 图 3 图 4 图 5 图 6 ...
Страница 80: ...80 PKG08861 UM733 1 DEUTSCH Abb 3 Abb 4 Abb 5 Abb 6 ...
Страница 94: ...94 PKG08861 UM733 1 ITALIANO ...
Страница 104: ...Starrett com ...