background image

7

Operating Instructions

 

Install the battery that comes with the instrument. See item “Installation/
Replacement  of  battery”.  Once  installed  the  caliper  will  come  on 
automatically.

 

To select the measurement unit needed, press the “mm/IN” button (13).

 

To move the slide (5) loosen the slide locking screw (8). Use the thumb 
roll (11) to move the slide.

 

Check to see if the buttons “zero” (12) and “mm/IN” (13) respond well. 
Digital display must be stable and display at any position within the whole 
range. 

 

Clean the measuring faces with a lint-free cloth or chamois.

 

External measurement:

 Close the jaws and press the “zero” button (12) to 

reset the display to zero (fig. 1A ). Put the part to be measured as close as 
possible to the scale (6)  (fig. 1B) and adjust the measuring faces (1) to the 
part surface (fig. 1C). Do not apply excessive pressure when measuring 
(fig. 2).

 

Internal measurement:

 Close the jaws and press the “zero” button (12) to 

reset the display to zero (fig. 3A). Introduce the internal measuring jaws (2) 
as deep as possible inside the part and adjust the measuring faces to the 
part surface (fig. 3B). To obtain the measurement of an internal diameter 
see fig. 3C and to obtain the measurement value inside a slot see fig. 3D.

 

Depth  measurement:

  Close  the  jaws  and  press  the  “zero”  button  (12) 

to reset the display to zero (fig. 4A). Keep the depth measuring rod (3) 
perpendicularly to the part bottom (fig. 4B & 4C).

 

Cam  measurement

:  Close  the  jaws  and  press  the  “zero”  button  (12)  to 

reset the display to zero (fig. 5A). Open the jaws in a little bigger dimension 
than the cam, place the scribing face (4) against the bottom of the cam and 
approach the scribing reference face (10) until the part surface. The cam 
value will be shown on the display (fig. 5B).

www.

.com

[email protected]

1.800.561.8187

Содержание 799A Series

Страница 1: ...1 www com information itm co 00 561 8187 ...

Страница 2: ... Asta di misurazione della profondità Tiefenmessstange 深度测杆 Haste para medição de profundidade 4 Scribing face Superficie de trazado Face de traçage Superficie di tracciatura Anreißfläche 划线面 Superfície para traçagem 5 Slide Cursor Coulisse Corsoio Schieber 滑动读数头 Cursor 6 Graduated scale Escala graduada Barre graduée Scala graduata Schiene mit Strichskala 刻度尺身 Escala graduada 7 LCD display Visuali...

Страница 3: ...Zoll mm 公 英制 转换键 Botão de conversão polegada milímetro mm IN 14 Power on off button Botón encendido apagado on off Bouton marche arrêt on off Pulsante di alimentazione on off An Aus Taste on off 开 关 键 Botão liga desliga on off NOTE The pictures shown are illustrative the components may vary according to the model Obs Las figuras aquí presentadas son ilustrativas los componentes pueden cambiar conf...

Страница 4: ...cation when the least significant digit changes by one step VIM 2 Accuracy Closeness of the agreement between the result of a measurement and a true value of the measurand VIM P S VIM International Metrology Vocabulary Set zero at any position of the slide Power on off button Automatic on when moving the slide Automatic off after seven minutes of non use to save the battery life Off at any positio...

Страница 5: ... may damage the circuitry of the caliper Do not disassemble the caliper Do not move the slide too fast over 1 5m s or 60 inches s this can cause measurement errors Flashing of digits shows a weak battery and that it must be replaced Do not wash or immerse the caliper in any liquid When cleaning the caliper use a fabric lightly wet in isopropyl alcohol Keep the caliper clean and dry Avoid handling ...

Страница 6: ...ace fig 1C Do not apply excessive pressure when measuring fig 2 Internal measurement Close the jaws and press the zero button 12 to reset the display to zero fig 3A Introduce the internal measuring jaws 2 as deep as possible inside the part and adjust the measuring faces to the part surface fig 3B To obtain the measurement of an internal diameter see fig 3C and to obtain the measurement value insi...

Страница 7: ...sing one of the holes or one of the pins as a reference press the zero button 12 to reset the display to zero fig 8A 9A After that start measuring the center to center value will be shown on the display fig 8B 9B Measurement of wall thickness Using a part as reference place the caliper and press the zero button 12 to reset the display to zero fig 10A Next measure the depth using the depth measurin...

Страница 8: ...not change when the slide is moved Trouble with the electronic circuitry Remove the battery and reinstall after 30 seconds Less accurate than specified or electronic system failure Contamination on the sensor Clean the scale If the problem remains contact your distributor No display on LCD 1 Battery in poor contact 2 Low battery 1 Remove the battery cover and reseat the battery to insure a good co...

Страница 9: ...46 www com information itm co 00 561 8187 ...

Страница 10: ...47 www com information itm co 00 561 8187 ...

Отзывы: