Regulatory Information | 43
42 | Regulatory Information
WIRELESS TECHNICAL DESCRIPTION
Your hearing aids contain a radio transceiver utilizing Bluetooth
®
Low Energy
wireless technology operating in the 2.4-2.4835 GHz frequency band with a
maximum effective radiated power of -6.9 dBm with transmission modulation
type of 814KFXD. The receiver section of the radio has a bandwidth of 1.5 MHz.
This hearing aid model has been tested to, and has passed, the following
emissions and immunity tests:
• IEC 60601-1-2 radiated emissions requirements for a Group 1 Class B device as
stated in CISPR 11.
• RF radiated immunity at a field level of 10 V/m between 80 MHz and 2.7 GHz
as well as higher field levels from communication devices as stated in Table 9
of IEC 60601-1-2.
• Immunity to power frequency magnetic fields at a field level of 30 A/m.
• Immunity to ESD levels of +/- 8 kV conducted discharge and +/- 15 kV
air discharge.
Hereby, Starkey
®
Hearing Technologies declares that the products listed above
are in compliance with the essential requirements and other relevant provisions of
Directive 2014/53/EU. A copy of the Declaration of Conformity can be obtained
from the addresses on the following page or docs.starkeyhearingtechnologies.com
WIRELESS NOTICES
FCC ID: EOA-24HALO2R312
IC: 6903A-24HALO2R312
FCC ID: EOA-24HALO2R13
IC: 6903A-24HALO2R13
FCC NOTICE
This device complies with part 15 of the FCC rules and with ISED Canada
license-exempt RSS standards. Operation is subject to the following two
conditions: (1) This instrument may not cause harmful interference, and (2) this
instrument must accept any interference received, including interference that
may cause undesired operation of the instrument.
NOTE:
The manufacturer is not responsible for any radio or TV interference
caused by unauthorized modifications to this equipment. Such modifications
could void the user’s authority to operate the equipment.
Starkey Hearing Technologies
6700 Washington Ave. South
Eden Prairie, MN 55344 USA
Wm. F. Austin House, Bramhall Technology Park
Pepper Road, Hazel Grove, Stockport SK7 5BX
United Kingdom
Waste from electronic
equipment must be handled
according to local regulations
Consult Operations Manual
Starkey Labs Canada Co.
2476 Argentia Road, Suite 301
Mississauga, ON L5N 6M1