Limited Warranty

StarDental

®

 Concentrix

®

 III

 and 

Concentrix

®

 IV 

Handpieces are warranted against defects in material and workmanship for 

6 months from date of purchase.  The Manual Chuck Turbine is warranted for 3 months. O-Rings and Seals are specifically 

excluded from this warranty.
This warranty is subject to the following conditions:
The product must not be abused and must have been operated in accordance with procedures detailed in this manual.
Any replacement parts must be genuine StarDental

®

 parts.

StarDental

®

 will, without charge, repair or replace (at its sole discretion) any defective part(s) covered in this warranty, 

providing above conditions are met and the product is returned through an authorized StarDental

®

 dealer.  You must 

have a copy of your original invoice to qualify.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE WARRANTIES 

OF MERCHANTABILITY AND FITNESS FOR USE.

 

Problem:  LACK OF POWER

 

Cause: 

  Air pressure

 

Correction: 

  Verify proper handpiece air pressure. (Refer to 

Specifications

.)

 

Cause: 

  Crimped tubing

 

Correction: 

  Straighten the tubing inside and outside delivery head.

 

Cause: 

  Debris buildup in handpiece

 

Correction: 

  Clean inside of housing and end cap with cotton swab and alcohol. Before installing turbine and end

 

  cap, lubricate the O-rings with 

DentaLube II

 using the soft end of a cotton swab.

 

Problem:  WATER OR AIR LEAKAGE

 

Cause: 

  Brittle handpiece rear seal

 

Correction: 

  Refer to 

Rear Seal Replacement

.

 

Cause: 

  Handpiece not threaded fully onto tubing

 

Correction: 

  Refer to appropriate 

Handpiece Installation

.

 

Problem:  IMPROPER SPRAY

 

Cause: 

  Clogged filters, tubes or water ports

 

Correction: 

  Inspect water filter in delivery head. 

 

  Disconnect handpiece. Depress foot control; verify that water exits 

 

  tubing. Insert cleaning wire provided, 

StarDental

 Part #256719, into spray tube near handpiece head

 

  to clear debris. Remove wire; aerate spray tubes with syringe air.

Manual Turbine Bur Changing System

1.  Pull handle back on manual bur tool, 

StarDental

 

Part #203108, and slide open end of tool onto  

square nut  as shown; release spring loaded handle  

and seat shaft of tool into the opening in the rear of  

turbine housing.

2.  To unlock chuck, turn manual bur tool handle  

counterclockwise.

3.  After inserting a 1.6 mm diameter friction grip bur  

fully into chuck, turn handle of tool clockwise to  

tighten chuck.

1.  Grasp the handpiece in front of the fixed connection 

between the thumb and index finger as shown.

2.  Slide tubing coupler until the end of the tubing 

connector is exposed; hold in hand as shown with 
fingers securing the tubing.

3.  Align and insert handpiece fixed connection tubes into 

the tubing connector holes.

4.  While applying slight pressure to hold fittings in the 

connector holes, slide coupler forward and engage 
threads. Continue threading coupler clockwise until 
hand tight against the handpiece.

Instruction Manual

Concentrix

®

 III and Concentrix

®

 IV

High Speed Handpiece

      

Do Not Use for Oral Surgery  

    

Contents

• Specifications
• Handpiece Installation
• Handpiece Operation
• Safety Rules

Note:

 

Follow the sterilization guidelines as stated by the manufacturer of your sterilizer.

• Cleaning and Lubricating
• Troubleshooting
• Turbine Replacement

• Rear Seal Replacement
• Replacement Part Numbers

• Warranty

Specifications

Concentrix

®

III 

High-Speed Handpiece, 

StarDental

®

 Part #263520 and 

Concentrix

®

IV 

High-Speed Handpiece, 

StarDental

®

 Part 

#262399

 

are conventional non-swivel high-speed handpieces.  The 

Concentrix III

 connects to either

 

 2-Line or 2/3-Line ISO 

tubing. The 

Concentrix IV

 connects to 4-Line ISO tubing.

 

Type:  Stainless steel high-speed handpiece with a replaceable manual chuck turbine

 

Operating Air Pressure:  30 to 32 psi (2.11 to 2.25 kgs/cm

2

)

 

Air Consumption:  1.9 SCFM (896.8 cms

3

/sec.) @ 30 psi (2.11 kgs/cm

2

)

 

No Load Speed:  300,000 rpm

 

Connection: 

Concentrix III: 

2- or 2/3-line fixed ISO configuration

 

 

Concentrix IV: 

4-line fixed ISO configuration

 

Power:  12 watts

 

Weight:  38.5g - 

Concentrix III, StarDental

®

 Part #263520

 

  40g - 

Concentrix IV

StarDental

®

 Part #262399

 

Sterilizable:  Steam Sterilization at 121° C or 135° C

Handpiece Installation

Note: 

Filter air and water supplies. Soften and deionize water. Be sure filters operate efficiently

To ensure efficient spray  

pattern, it is recommended to use a 12 micron water fliter.

Manual Bur Changing System

Maximum

24.0 mm

OVERALL LENGTH

Minimum

 

10.8 mm

Shank Length

Use standard length concentric diamond and carbide burs with 1.6mm nominal diameter friction grip shanks designed for 
high speed handpieces. Every 

StarDental

 Diamond Bur meets specifications below for recommended lengths. Contact your 

dealer for a complete list of 

StarDental

 Diamond and/or Carbide Burs. 

Star Dental

 Manual turbines have the following bur requirements:  

Use only burs with hardened, tempered steel shanks or carbide shanks. Every 

StarDental

 

Bur maintains required hardness. Before use, verify concentric rotation by operating 
handpiece outside the oral cavity in a safe place.

Note:

 Do not use burs that produce excessive noise or do not run concentrically. Do 

not use burs that are modified, eccentrically designed, or are bent or 

soft

 (non-

heat treated) shanks.

Handpiece Operation

WARNING:

 

Use burs with an overall length of less than 24 mm and a shank length of at least 10.8 mm.

4.  Disengage tool and tug on bur to verify that it is completely engaged and holding securely. If chuck does not hold bur 

securely, clean or replace chuck.

CAUTION:

 

Be careful to keep bur tool aligned with handpiece. If tool is side-loaded while engaged in chuck, tool may 

be damaged.

REMOVAL of Bur from the Manual Chuck

1.  Use manual bur tool and turn counterclockwise one full turn, 360° rotation. 

2.  Pull bur straight out.

 

Safety Rules

1.  Always tug on bur after chuck is loaded to confirm that the bur is secured.
2.  Before use, verify concentric rotation by operating the handpiece outside of the oral cavity in a safe place. Do not use 

burs that do not run concentrically or produce excessive noise.

3.  Never operate handpiece without a bur installed.
4.  Use burs with an overall length of less than 24 mm.  Shank length must be at least 10.8 mm.
5.  Use only recommended burs to prevent damage to handpiece cartridge and/or chuck, and in some cases to prevent burs 

from improper or incorrect placement.  An improperly clamped bur may dislodge during use, causing injury to you or 
your patient. A sudden, high-pitched sound or increased vibration while handpiece is operating alerts the operator to 
the possibility that the chuck is not clamped sufficiently to retain the bur or a bent bur is inserted. 

PLEASE OBSERVE SAFETY RULES

Cleaning and Lubricating

The 

Concentrix III

 and 

Concentrix IV 

Handpieces’  bearings are lubricated and sealed to keep the dirt out, therefore never 

clean ultrasonically or immerse in solvent that would dilute the lubricant.

Cleaning the Chuck - clean prior to sterilization (after each patient)

Clean the chuck  once a day or as required. Remove bur from chuck; place 2 drops 

of 

DentaLube

 

II

StarDental

 

Part  #262539,  into  chucking  mechanism. Aerate  with 

syringe air.

Monthly Cleaning of the Chuck

 1.  Loosen chuck with bur tool; remove bur.

 2.  Turn bur tool knob counterclockwise until chuck emerges from end cap.

 3.  Remove bur tool; turn chuck with fingers to free.

 4.  Clean chuck with alcohol 

 5.  Run paper through 3 slots of chuck; remove debris.

 6.  Reinstall chuck by reversing steps 1-3.

LUBRICATING the Turbine

Note:

 

Lubricate handpiece

 

prior

 to each autoclave sterilization procedure. The use of the 

 

grease syringe is no longer necessary. Use 

DentaLube II

 

as described in this section.

 

After  each  patient,  flush  with  air  and  water  by  running  handpiece  for  20  seconds

.

1.  Disconnect handpiece from tubing.

2.  Insert the red tip applicator, 

StarDental

 Part #202954, into the 

DentaLube II

 

bottle cap.

3.  Identify drive air tube  at rear of handpiece and place three drops of  

DentaLube II, StarDental

 Part #262539.  (See illustration for more detail).

4.  With bur in place connect handpiece to tubing, operate for 20 seconds to  

expel excess lubricant.

5.  Disconnect handpiece from tubing.

6.  Clean all external surfaces using a small toothbrush with isopropyl alcohol.

7.  Remove Bur.

8.  Sterilize - See Sterilization

9.  Remove handpiece from sterilizer after drying cycle end and allow to cool in bag.

Troubleshooting

Turbine & O-Ring Replacement

1.  Remove end cap by unthreading counterclockwise.
2.  Push cartridge out back of handpiece.
3.  Remove o-rings from housing using an explorer.
4.  Clean housing and O-ring grooves using a cotton swab soaked 

in isopropyl alcohol.  All debris and pieces of O-ring must be 
removed from the grooves.

5.  Using the set of tools 

StarDental

 Part #263568, Insert the 

insertion cap all the way down until it stops as shown. 

(Detail 1) 

6.  Place O-ring inside housing, and guide it to the groove. 

 

 

Make sure O-ring is past exhaust hole, or the O-ring could be 

cut when trying to insert base. (Detail 2)

7.  Insert the base gently until it stops on housing  as shown. 

(Detail 3 & 4)

8.  While holding the base against housing, push on the insertion cap and O-ring 

until it is against the base.  Pull the base out gently and check the O-ring,  
O-ring should be seated in lower groove. Failure to insert O-ring in groove after 
two attempts may be the result of debris in groove. Clean groove and follow 
steps 5 through 8.  

(Detail 4 & 5)

9.  Push the tool flush with the lower edge of the upper groove and guide new O-ring as best as possible to the groove.
10.  While holding the insertion cap in place, push base gently into housing driving O-ring into groove. 

(Detail 6)

11.  Remove base verify that the O-ring is in the groove, then push insertion cap out. 

(Detail 7)

12.  Install 

new 

turbine. Thread end cap on. 

(Detail 8)

Rear Seal Replacement

DentalEZ Dental Products (GB) Ltd. 
Cleveland  Way,  Hemel Hempstead, 
HP2 7DY, ENGLAND

63518 E  Date of Issue 8-10-2010

Division of DentalEZ Group  

Lancaster, PA 17601-5891   U.S.A.

Printed in the U.S.A.

©

2010 DentalEZ Group

Seals wear with repeated use. Once a year or as needed, replace the rear seal in the handpiece: for the 2/3-line handpiece re-
place rear seal with 

StarDental

 Part #203103, and for the 4-line handpiece replace rear seal with 

StarDental

 Part #202897.

Replacement Part Numbers

Always insist on StarDental parts for best performance.

Cleaning Wire (Package of 2, 0.008 Gauge) ................256719
Rear Seal, 2/3-Line (Pkg. of 3) .....................................203103
Rear Seal, 4-Line (Pkg. of 3) ........................................202897
Turbine and O-Rings Replacement Kit .......................203111
Chuck .........................................................................203098

Bur Tool ......................................................................203108
Air Pressure Gauge for 4-Line  ....................................262591
End Cap ..................................................................... 203099

DentaLube

 

II

    ...........................................................262539

Set of Tools (for mounting the O-rings) .....................263568

Description

Item

Description

Item

Handle

Manual

Bur Tool

Slide Open End of 

Tool Down

Bur

Chuck

Chuck Slot

Clean debris from 

inside of housing

Clean debris from

inside of Spray Tube

Product complies

with MDD 93/42/EEC

0344

Rear Seal

2/3-Line

2

3

Base

O-ring

1

Insertion 

cap

O-ring

Grooves

4

6

5

7

8

E n d 

Cap

Turbine

Air and Water Adjustment

1.  For 4-Line: using an air pressure gauge, 

StarDental

 Part #262591, set pressure to 30 psi (2.1 kg/cm

2

).

 

For 2/3-Line: if a gauge is available, set pressure to 30 psi (2.1 kg/cm

2

); if a gauge is not available, set pressure at 

delivery head allowing 1 psi (.0703 kg/cm

2

) pressure loss for every foot (30 cms) of tubing. Example:  

 

 

 

 

for 5 feet (1.52 m.) tubing: set control at 35 psi. (2.46 kgs/cm

2

)

 

 

 

or

 

for 12 feet (3.65 m.) tubing: set control at 42 psi.  (2.95 kgs/cm

2

)

Note: 

For maximum turbine life, do not exceed recommended pressure.

2.  Water Pressure –  Adjust water volume at the delivery head to obtain a satisfactory spray mist.

Concentrix IV 

Handpiece

Tubing

4-line Tubing Connector

Tubing Coupler

Concentrix III

Handpiece

Tubing

2/3-line Tubing Connector

Tubing Coupler

Rear Seal

4-Line

Sterilization

Handpiece, chuck and burs must be cleaned and sterilized after each patient.

Note:

 

To Prevent Damage To the Bur, Use an Autoclavable Bur Block.

wrapped

 handpiece or bur  should be steam sterilized at 132° C for a minimum of 10 minutes or at 121° C for a 

minimum of 50 minutes. Use paper or plastic bags only.  Do not exceed 135° C.
An 

unwrapped

 handpiece or bur should be steam sterilized at 132° C for a minimum of 8 minutes or at 121° C for a 

minimum of 30 minutes. Do not exceed 135° C.

Note:

 

An unwrapped handpiece or bur needs to be used immediately or sterility is not maintained.

Do not exceed the sterilizer manufacturer’s guidelines for load capacity.

CAUTION: 

Use steam sterilization exclusively for complete sterility. Never dry heat sterilize, ultrasonically 

clean, or expose handpiece, bur or swivel to chemical disinfection or cold sterilization by:

• Immersion techniques or Surgical milk
• Surface wipes, Hand soaps, or Detergents
• Surface sprays
• Sterilization bags containing disinfectants

CLEANING Burs

 NOTE: Burs 

Must

 Be Cleaned and Sterilized After Each Patient.

Cleaning Diamond Burs:

     

- After each use, clean the diamond bur by running it against a wet diamond cleaning stone.

     

- Then, the diamond bur should be cleaned by being held in a solution of either enzymatic presoak or water.

     

- Rinse the bur thoroughly and allow it to dry.

     

- Sterilize – See Sterilization

Cleaning Carbide Burs:

     

- After each use, presoak the bur in solution of either enzymatic presoak or water.

     

- Rinse the bur thoroughly and allow it to dry.

     

- The bur may be cleaned with an autoclavable bur brush to remove any resistant debris from the bur.

     

- Sterilize – See Sterilization

Red Tip 

Applicator

Drive Air 

Tube 2/3-

Line

Drive Air Tube

4-Line

M A K E S   Y O U R   P R A C T I C E   P E R F E C T

®

Sterilizable up to 135°C

135°C