1.2 Front-Panel Operation
1 – 3
Figure 1.1: The SIM984 front and rear panels.
1.2.3
Bandwidth
User bandwidth settings of DC–100 Hz, DC–10 kHz, and DC–1 MHz
are with the left and right [Bandwidth] buttons.
1.2.4
Output
The amplified, band-limited signal appears at the (ground-referenced)
BNC connectors on the front and rear panels. Each output connec-
tion is through a 50
Ω
resistor. For normal operation, the user should
not need to terminate the output. If a 50
Ω
termination is applied,
the output signal will be divided in half.
If the output signal exceeds
±
10 V, the output overload detection is
activated. This is indicated by the red
OVLD
near the top of the
“OUTPUT” block on the front panel.
The SIM984 output o
ff
set voltage may be trimmed by adjusting the
SIM984 Isolation Amplifier