ST&G 1322 AL Скачать руководство пользователя страница 8

Note: 

4-bar knees inherently are very stable due to the geometry built into each 

design. This is commonly referred to as the Instant Knee Center (IKC). The IKC 
point when doing bench alignment, will fall behind the traditional TKA line that we 
will reference. (Fig. 2a,2b) Tg line in Fig. 3 is ideal placement, but in certain 
instances, it may be necessary to accommodate placement anteriorly (0 to 
10mm). The Tg line is referencing a moving A/P weight bearing line, so it could 
be slightly anterior or neutral.

a) With prosthesis assembled, taking into account hip flexion contractures, 
abduction, Line Of Progression, and toe out (Fig.2a), the TKA plumb line should 
pass through the knee center (center ot proximal/anterior pivot Fig.2a, 2b) and in 
front of the K point (IKC). 

NOTE: Take into account shoe heel height, and add 3mm safety factor.

b) Ideally, the pylon connecting the knee and foot should end up vertical. There 
may be a variance due to the foot alignment recommendations. In this case, the 
maximum anterior tilt of the pylon should not exceed 3 degrees, and it may be 
necessary to utilize 1222T offset tube clamp adapter. 

c) With prosthesis donned, the weight line should pass through the centerline of 
the knee in the Coronal or M/L plane (Fig. 2c). Excessive outset or inset will put 
undue stress on the knee joint.

d) With prosthesis donned, the weight line for Sagittal or A/P plane should have 
the plumb line passing ideally through the knee center (proximal anterior pivot), 
and be perpendicular to the ground. (Fig. 2a, 2b)

7.1     BENCH ALIGNMENT:

It is not recommended to have alignment posterior to the reference line, as 
it could cause knee instability!

Содержание 1322 AL

Страница 1: ...Instructions for 1322 AP AL Polycentric Pneumatic 4 Bar Knee ST G USA Corp Phone 714 524 0663 2691 Saturn St Fax 714 364 8113 Brea CA 92821 www stngco com...

Страница 2: ...o 125kg 275lbs Contra indications Residual muscular weakness contractures or proprioceptive dysfunction including poor balance Contra lateral joint instabilities or pathology Complicated conditions in...

Страница 3: ...ral View of Knee Unit 2 Construction Principal Parts Frame Aluminum Alloy Brass Stainless Steel Steel Knee Aluminum Alloy Stainless Steel Knee control Various materials principally Aluminum Alloy Stai...

Страница 4: ......

Страница 5: ...fect the stability of the knee The user should be advised to contact their Clinician Practitioner if their condition changes 4 Safety Information 3 Function Adjustable spring extension assist Independ...

Страница 6: ...orinated water it should be rinsed with fresh water and dried 6 Limitations on use Intended Life Service life of the product is covered by the warranty period 2 years This product is recommended for u...

Страница 7: ...7 Alignment and Set Up Users be aware of potential finger trap hazard Fig 2a Fig 2b Fig 2c...

Страница 8: ...2a 2b and in front of the K point IKC NOTE Take into account shoe heel height and add 3mm safety factor b Ideally the pylon connecting the knee and foot should end up vertical There may be a variance...

Страница 9: ...or higher or lower walking speeds Swing Phase Control Adjustment It is advisable to adjust flexion before extension for optimum walking symmetry If needed please follow directions below Ensure full kn...

Страница 10: ...m driver Note Mark indicate pyramid orientation Loosen set screw using 2 5mm driver Use 6mm driver and turn clockwise to increase extension assist Turn screw anti clockwise to reduce the extension ass...

Страница 11: ...of knee head Apply glue to new ones and insert back into knee head Use a small screw driver to pick out the extension stop rubber bumper on knee level adjusting screw Insert new one into slot 9 Mainte...

Страница 12: ...vity K2 to K3 Maximum User Weight 125kg 275lbs Maximum flexion angle 135 degrees Proximal Alignment attachment Male Pyramid or Lotus Adapter Distal Alignment attachment Tube Clamp Tube clamp torque se...

Страница 13: ...anty Warranted for 2 years from the date of invoice by ST G The user should be aware that changes or modifications not approved will void the warranty 11 Liability The manufacturer recommends using th...

Страница 14: ...ST G USA Corporation www stngco com e mail info stngco com 2691 Saturn Street Brea CA 92821 USA Tel 1 714 524 0663 Fax 1 714 364 8113 1322IFU Rev B 01 16 18...

Отзывы: