General Specifications
1. Maximum voltage including transient over-voltage between any terminals and grounding: 500V RMS.
2. Fused protection for VΩ mA input terminal: 315 mA, 250V fast type, 5 x 20 mm
3. 10A Terminal: Un-fused.
4. Range: Manual ranging.
5. Maximum Display: 1999.
6. Measurement Speed: Updates 2-3 times/second.
7. Temperature:
Operating: 0~40°C (32°~104° F).
Storage:-10°~50° C (14°~122°F).
8. Relative Humidity: ≤75% @ 0°~30° C; ≤ 50% @ 31°~40° C
9. Altitude: Operating: 2,000m; Storage: 10,000m.
10. Battery Type: One 9V Battery, NEDA 1604, 6F22, or 006P.
11. Battery Deficiency: Display:
12. Negative reading: Display: —
13. Overload: Display: 1
14. Accuracy: ± (a% r b digits, see charts below), guarantee for 1 year.
Accuracy Specifications
7
Note:
Input impedance: 10 MΩ
Note:
1
. Input impedance: approx. 5 MΩ
2. Displays effective value of sine wave (mean value response).
3. Frequency response: 40 Hz~400 Hz.
Note:
At
10A range: For continuous measurement ≤ 10
seconds and interval less than 15 minutes.
DC Current
AC Voltage
DC Voltage
Resistance
Note:
1. No overload protection.
2. Make sure voltage output of calibrated equipment
level is less than 10V to avoid damage to the meter.