Page
ST312B User Handbook
8
.
°C
°C/hr
hr.min
The next push of the
key displays the
soak time in the range 00.00 to 99.59 (99
hours 59 mins). This can be altered with
the
&
keys. The SOAK TIME lamp
on the mimic display will flash.
Note: to exit programming without cycling through all of the above steps wait 10
seconds without pressing any keys
-
the ST312B will revert to the idle display.
Alternatively press the START/STOP key to exit programming and to begin firing
immediately.
Pr
ogram
ming
A further push of the
key completes data
entry and no lamps on the mimic display will
flash.
°C
°C/hr
hr.min
Data entry is also completed if
is selected
with the
key for either ramp rate 1 or ramp
rate 2.
Note: the
key can be used to reverse through the programming steps to correct
errors or to exit programming mode.
Note: available ramp rate displays are:
,
…
&
.
If
is
shown but another segment is required then push the
key to obtain the
required ramp rate (in the range 1°C/hr to 999°C/hr). If full power is required
then push the
key until
is diplayed.